Only Zero Calibration (355 I.S.):1. From the <strong>Weigh</strong> Mode, press [1] [0] [0] [SELECT].2. Press [5] [4] [3] [2] ]1] [ENTER].3. Press [SELECT] [SELECT] [SELECT].4. Remove any load on the scale.5. Press [ENTER] to establish zero.6. Pause for motion delay.7. Press [ENTER] to accept calibration.8. Press [ENTER] to save calibration.9. Press [ENTER] to exit calibration.5.2.6 Reset CalibrationCal Reset may be necessary when an over-load or under-load condition exists,preventing the completion of the calibration process. Calibration Reset adjusts the zeroand gain factors of the A/D amplifier to factory default values for maximum sensitivity.After performing a calibration reset, a complete recalibration is required. The effects ofa calibration reset do not take effect until the Model 350/355 I.S. is recalibrated andcalibration information has been saved.If Code 02 (under-load) or Code 03 (over-load) is displayed during calibration, press[ZERO] or [CLR] to perform a calibration reset.Reset Calibration Gain Factors (350 I.S.):1. Press [ZERO] + [SELECT].2. Press [ZERO] [PRINT] [UNITS] [TARE].3. Remove any load on the scale.4. Press [TARE] to establish zero.5. Place a test weight on scale.6. Press [ZERO].7. Remove any load on the scale.8. Press [TARE] to establish zero.9. Pause for motion delay.10. Place a test weight on scale.11. Enter the test weight value with the [PRINT] and [UNITS] keys.12. Press [TARE] to establish span.122Model 350 I.S. and Model 355 I.S. <strong>Service</strong> <strong>Manual</strong>

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!