First Zero Calibration Method (355 I.S.):1. From the <strong>Weigh</strong> Mode key in [1] [0] [0] [SELECT].2. Key in [5] [4] [3] [2] ]1] [ENTER].3. Remove any load on the scale.4. Press [ENTER] to establish zero.5. Pause for motion delay.6. Place a test weight on scale.7. Enter the value of the test weight with the numeric keys.8. Press [ENTER] to establish span.9. Pause for motion delay.10. Press [ENTER] to accept calibration.11. Press [ENTER] to save calibration.12. Press [ENTER] to exit calibration.13. Remove the calibration weight.5.2.3 Last ZeroThe Last Zero procedure allows recalibration of the weighing device using an existingtest load. This is especially beneficial when checking high capacity applications suchas tank weighing to minimize the task of placing and removing test weights.Establish gross zero before entering setup or calibration!Last Zero Calibration with <strong>Weigh</strong>t Already Applied (350 I.S.):1. Remove any load on the scale.2. Press [ZERO] to zero the scale.3. Apply a test weight to verify calibration.4. Press [ZERO] + [SELECT].5. Press [ZERO] [PRINT] [UNITS] [TARE].6. Press [SELECT].7. Press [TARE] to use last zero.8. Enter the value of the test weight with the [PRINT] and [UNTIS] keys.9. Press [TARE] to establish span.10. Pause for motion delay.11. Press [TARE] to accept calibration.Model 350 I.S. and Model 355 I.S. <strong>Service</strong> <strong>Manual</strong> 119

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!