Auto Hematology Analyzer User’s <strong>Manual</strong>1) Have calibration sample ready, press Aspiration key to start test.2) After test, calibration result will be shown in current test number column.3) Repeat test, the instrument allows 20 runs at most.4) Statistic results and new calibration coefficient will be shown in table below frame aftereach test automatically.5) Unsatisfactory test data rows can be deleted at any time.If deciding to use new calibration coefficient, click “Save” button, the system will save newcalibration coefficient.32