Auto Hematology Analyzer User’s <strong>Manual</strong>3 Sample test3.1 Preparation before startupBefore startup, the operator shall check by following steps to make sure system is ready:1. Check if the diluent, Lyse and cleanser meet the need of current test. Check if the wastebottle is full, tubing system is ok and without folded, all the connection is ok.2. Whether instrument power adapter is safely and stably connected to power socket.3. Check whether built-in or external printer paper is sufficiently installed in right position.3.2 StartupSwitch on the , indicator light is on, system begins to check <strong>com</strong>ponent status, transferparameter needed by front end, run blank tests, as shown in figure below:Fig.3-1 Startup programAfter startup self-checking, go to sample test screen, as shown below:Fig.3-2 Blank test resultBlank test only shows test results of WBC, RBC, HGB, HCT and PLT. The operator can runblank test at any time required. In sample test menu, click “Profile” button, after Data Editmenu pops, change sample no.. to 999, then click “Yes” button, return to sample test menu.Press “Aspiration Key” to start blank test. During test, the indicator light turns red, statuscolumn on top of screen will display test status.14