HP 8753ET and 8753ES Network Analyzers 30 kHz to 3 or 6 GHz

HP 8753ET and 8753ES Network Analyzers 30 kHz to 3 or 6 GHz HP 8753ET and 8753ES Network Analyzers 30 kHz to 3 or 6 GHz

System perf<strong>or</strong>mance summaries<strong>HP</strong> <strong>8753ES</strong> (50-ohm systems)3.5-mm test p<strong>or</strong>tsConfiguration<strong>Netw<strong>or</strong>k</strong> analyzerCalibration kitTest-p<strong>or</strong>t cablesSystem dynamic range<strong>30</strong> <strong>kHz</strong> <strong>to</strong> 50 <strong>kHz</strong> 70 dB 150 <strong>kHz</strong> <strong>to</strong> <strong>30</strong>0 <strong>kHz</strong> 90 dB 1<strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 16 MHz 100 dB16 MHz <strong>to</strong> 3 <strong>GHz</strong> 110 dBOption 014 108 dB3 <strong>GHz</strong> <strong>to</strong> 6 <strong>GHz</strong> 105 dBOption 014 103 dB<strong>HP</strong> <strong>8753ES</strong>St<strong>and</strong>ard <strong>and</strong> Option 006<strong>HP</strong> 85033D<strong>HP</strong> 11857DMeasurement p<strong>or</strong>t characteristicsFrequency rangeC<strong>or</strong>rected <strong>30</strong> <strong>kHz</strong>-<strong>30</strong>0 <strong>kHz</strong> 2 <strong>30</strong>0 <strong>kHz</strong>-1.3 <strong>GHz</strong> 1.3 <strong>GHz</strong>-3 <strong>GHz</strong> 3 <strong>GHz</strong>-6 <strong>GHz</strong>Directivity 49 dB 46 dB 44 dB 38 dBSource match 49 dB 44 dB 41 dB 37 dBLoad match 49 dB 46 dB 44 dB 38 dBReflection ±(0.010 dB ±(0.005 dB ±(0.007 dB ±(0.009 dBtracking +0.02/˚C) +0.01/˚C) +0.02/˚C) +0.03/˚C)Transmission ±(0.016 dB ±(0.014 dB ±(0.022 dB ±(0.048 dBtracking +0.02/˚C) +0.01/˚C) +0.02/˚C) +0.03/˚C)Reflection uncertaintyTransmission uncertaintyMagnitudePhaseMagnitudePhase1. Typical below <strong>30</strong>0 <strong>kHz</strong>.2. Typical perf<strong>or</strong>mance.6


System perf<strong>or</strong>mance summaries<strong>HP</strong> <strong>8753ES</strong> (75-ohm systems)Type-N test p<strong>or</strong>tsConfiguration<strong>Netw<strong>or</strong>k</strong> analyzer <strong>HP</strong> <strong>8753ES</strong> Option 075Calibration kit <strong>HP</strong> 85036BTest-p<strong>or</strong>t cables <strong>HP</strong> 11857BSystem dynamic range<strong>30</strong> <strong>kHz</strong> <strong>to</strong> 50 <strong>kHz</strong> 68 dB 150 <strong>kHz</strong> <strong>to</strong> <strong>30</strong>0 <strong>kHz</strong> 90 dB 1<strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 16 MHz 96 dB16 MHz <strong>to</strong> 3 <strong>GHz</strong> 106 dBReflection uncertaintyMeasurement p<strong>or</strong>t characteristicsFrequency rangeC<strong>or</strong>rected <strong>30</strong> <strong>kHz</strong> - <strong>30</strong>0 <strong>kHz</strong> 2 <strong>30</strong>0 <strong>kHz</strong> - 1.3 <strong>GHz</strong> 1.3 <strong>GHz</strong> - 3 <strong>GHz</strong>Directivity 48 dB 48 dB 43 dBSource match 47 dB 41 dB 35 dBLoad match 48 dB 48 dB 43 dBReflection ±(0.004 dB ±(0.010 dB ±(0.019 dBtracking +0.02 dB/˚C) +0.01 dB/˚C) +0.02 dB/˚C)Transmission ±0.018 dB ±(0.016 dB ±(0.033 dBtracking +0.02 dB/˚C) +0.01 dB/˚C) +0.02 dB/˚C)Unc<strong>or</strong>rected 2 <strong>30</strong> <strong>kHz</strong> - <strong>30</strong>0 <strong>kHz</strong> <strong>30</strong>0 <strong>kHz</strong> - 1.3 <strong>GHz</strong> 1.3 <strong>GHz</strong> - 3 <strong>GHz</strong>Directivity 20 dB* 35 dB <strong>30</strong> dBSource match 16 dB** 16 dB 16 dBLoad match 15 dB** 18 dB 16 dBReflection ±2.5 dB ±1 dB ±1 dBtrackingTransmission ±2.5 dB ±1 dB ±1 dBtrackingCrosstalk 90 dB*** 100 dB 100 dB* 15 dB, <strong>30</strong> <strong>to</strong> 50 <strong>kHz</strong> **10 dB, <strong>30</strong> <strong>to</strong> 50 <strong>kHz</strong> *** 60 dB, <strong>30</strong> <strong>to</strong> 50 <strong>kHz</strong>Transmission uncertaintyMagnitudePhaseMagnitudePhase1. Typical below <strong>30</strong>0 <strong>kHz</strong>.2. Typical perf<strong>or</strong>mance.7


System perf<strong>or</strong>mance summaries<strong>HP</strong> <strong>8753ES</strong> (75-ohm systems)Type-F test p<strong>or</strong>tsConfiguration<strong>Netw<strong>or</strong>k</strong> analyzer <strong>HP</strong> 8753E Option 075Calibration kit <strong>HP</strong> 85039BTest-p<strong>or</strong>t cables <strong>HP</strong> 11857BSystem dynamic range<strong>30</strong> <strong>kHz</strong> <strong>to</strong> 50 <strong>kHz</strong> 68 dB 150 <strong>kHz</strong> <strong>to</strong> <strong>30</strong>0 <strong>kHz</strong> 90 dB 1<strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 16 MHz 96 dB16 MHz <strong>to</strong> 3 <strong>GHz</strong> 106 dBMeasurement p<strong>or</strong>t characteristicsData is shown f<strong>or</strong> a Type-F female reflection p<strong>or</strong>t <strong>and</strong> aType-F male transmission p<strong>or</strong>t.Frequency rangeC<strong>or</strong>rected 2 <strong>30</strong> <strong>kHz</strong>-<strong>30</strong>0 <strong>kHz</strong> <strong>30</strong>0 <strong>kHz</strong>-1.3 <strong>GHz</strong> 1.3 <strong>GHz</strong>-3 <strong>GHz</strong>Directivity 45 dB 45 dB 40 dBSource match 40 dB 40 dB <strong>30</strong> dBLoad match 45 dB 45 dB 40 dBReflection ±(0.060 dB ±(0.060 dB ±(0.024 dBtracking +0.02 dB/˚C) +0.01 dB/˚C) +0.02 dB/˚C)Transmission ±(0.033 dB ±(0.019 dB ±(0.057 dBtracking +0.02 dB/˚C) +0.01 dB/˚C) +0.02 dB/˚C)Reflection uncertaintyMagnitudePhaseTransmission uncertaintyMagnitudePhase1. Typical below <strong>30</strong>0 <strong>kHz</strong>.2. Typical perf<strong>or</strong>mance.8


System perf<strong>or</strong>mance summaries<strong>HP</strong> <strong>8753ET</strong> (50-ohm systems)Type-N test p<strong>or</strong>tsConfiguration<strong>Netw<strong>or</strong>k</strong> analyzer <strong>HP</strong> <strong>8753ET</strong> St<strong>and</strong>ard,Option 006 <strong>or</strong> Option 004Calibration kit <strong>HP</strong> 85032BTest p<strong>or</strong>t cable <strong>HP</strong> part number 8120-4781System dynamic range<strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 16 MHz 100 dB16 MHz <strong>to</strong> 3 <strong>GHz</strong> 110 dB3 <strong>GHz</strong> <strong>to</strong> 6 <strong>GHz</strong> 105 dBMeasurement p<strong>or</strong>t characteristicsFrequency rangeUnc<strong>or</strong>rected <strong>30</strong>0 <strong>kHz</strong>-1.3 <strong>GHz</strong> 1.3 <strong>GHz</strong>-3.0 <strong>GHz</strong> 3.0 <strong>GHz</strong>-6.0 <strong>GHz</strong>Directivity <strong>30</strong> dB 24 dB 19 dBSource match 4 25 dB 20 dB 14 dBLoad match 24 dB 19 dB 16 dBReflection tracking ±1.0 dB ±1.0 dB ±2.0 dBTransmission tracking ±1.5 dB ±1.5 dB ±2.5 dBCrosstalk 100 dB 100 dB 90 dBFrequency range (continued)C<strong>or</strong>rected <strong>30</strong>0 <strong>kHz</strong>-1.3 <strong>GHz</strong> 1.3 <strong>GHz</strong>-3.0 <strong>GHz</strong> 3.0 <strong>GHz</strong>-6.0 <strong>GHz</strong>Reflection measurements 1Directivity 50 dB 47 dB 40 dBSource match 42 dB 36 dB 31 dBLoad matchOne-p<strong>or</strong>t cal 24 dB 19 dB 16 dBEnhanced reflection cal 24 dB 19 dB 16 dBReflection tracking ±(0.009 dB ±(0.019 dB ±(0.07 dB+0.01 dB/˚C) +0.02 dB/˚C) +0.03 dB/˚C)Transmission measurements 2Source matchEnhanced-response cal 42 dB 36 dB 31 dBResponse-only cal 3 25 dB 20 dB 14 dBTransmission trackingEnhanced-response cal ±(0.006 dB ±(0.018 dB ±(0.054 dB+0.01 dB/˚C) +0.02 dB/˚C) +0.03 dB/˚C)Response-only cal ±(0.033 dB ±(0.1 dB ±(0.27 dB+0.01 dB/˚C) +0.02 dB/˚C) +0.03 dB/˚C)Transmission uncertainty:(enhanced-response calibration)Transmission uncertainty:(response calibration)Reflection uncertainty(one-p<strong>or</strong>t calibration)1. One-p<strong>or</strong>t <strong>or</strong> enhanced-response calibration.2. Enhanced-response <strong>or</strong> response-only calibration.3. 2 dB less with Option 004.4. Option 004 may degrade unc<strong>or</strong>rected source match as much as 2 dB.9


<strong>HP</strong> <strong>8753ES</strong> specificationsTest-p<strong>or</strong>t output characteristics 7Frequency characteristicsRange<strong>30</strong> <strong>kHz</strong> <strong>to</strong> 3 <strong>GHz</strong>Option 006 <strong>30</strong> <strong>kHz</strong> <strong>to</strong> 6 <strong>GHz</strong>Resolution 1 HzStability 9 ±7.5 ppm (0˚ <strong>to</strong> 55 ˚C)±3 ppm/yearAccuracy ±10 ppm at 25 ˚C ± 5 ˚COption 1D5 ±0.05 ppm (0˚ <strong>to</strong> 55 ˚C),±0.5 ppm/yearPower range 2 -85 <strong>to</strong> +10 dBmOption 075/014 -85 <strong>to</strong> +8 dBmPower sweep range 25 dB (typically 31 dB)Resolution 0.01 dBLevel accuracy 1,2,5 ±1.0 dBLevel linearity 1,2,5 -15 dBm <strong>to</strong> +5 dBm ±0.2 dB5 dBm <strong>to</strong> 10 dBm 6 ±0.5 dBImpedance 9 50 Ω (nominal)<strong>30</strong> <strong>kHz</strong> <strong>to</strong> 3 <strong>GHz</strong> >16 dB RL (14 dB RL (16 dB RL (


Test p<strong>or</strong>t input characteristics (continued)External source mode 1 (CW time sweep only)Frequency range <strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 6 <strong>GHz</strong>R-channel input requirements (typical)Power level 0 <strong>to</strong> -25 dBmSpectral purityMaximum


<strong>HP</strong> <strong>8753ES</strong> Option 011 specificationsTest p<strong>or</strong>t output characteristicsFrequency characteristicsRange <strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 3 <strong>GHz</strong>Option 006 <strong>30</strong> <strong>kHz</strong> <strong>to</strong> 6 <strong>GHz</strong>Resolution 1 HzStability typically ±7.5 ppm 0˚ <strong>to</strong> 55 ˚Ctypically ±3 ppm/yearOption 1D5 typically ±0.05 ppm 0˚ <strong>to</strong> 55 ˚Ctypically ±0.5 ppm/yearAccuracy ±10 ppm at 25 ˚C ± 5 ˚CPower range -5 <strong>to</strong> +20 dBmOption 006 -5 <strong>to</strong> +18 dBmResolution 0.01 dBLevel accuracy 1,2 ±1.0 dBLevel linearity 1,2,5 ±0.25 dB, -5 <strong>to</strong> +15 dBm±0.5 dB, +15 <strong>to</strong> +20 dBmImpedance 50 Ω nominal<strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 3 <strong>GHz</strong> 2 >16 dB RL (14 dB RL (


<strong>HP</strong> <strong>8753ES</strong> Option 011 specificationsTest p<strong>or</strong>t input characteristics (continued)Frequency offset mode 3Frequency range <strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 3 <strong>GHz</strong>Option 006 <strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 6 <strong>GHz</strong>R-channel input requirementsPower level<strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 3 <strong>GHz</strong> 0 <strong>to</strong> -35 dBm3 <strong>GHz</strong> <strong>to</strong> 6 <strong>GHz</strong> 0 <strong>to</strong> -<strong>30</strong> dBmLO spectral purityMaximum spurious


<strong>HP</strong> <strong>8753ET</strong> specificationsTest-p<strong>or</strong>t output characteristicsFrequency characteristicsRange <strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 3 <strong>GHz</strong>Option 006 <strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 6 <strong>GHz</strong>Resolution 1 HzStability 4 ±7.5 ppm (0˚ <strong>to</strong> 55 ˚C)±3 ppm/yearOption 1D5 ±0.05 ppm (0˚ <strong>to</strong> 55 ˚C)±0.5 ppm/yearAccuracy ±10 ppm at 25 ˚C ±5 ˚CPower range: -20 <strong>to</strong> +5 dBm-85 <strong>to</strong> +10 dBm (with Option 004)Resolution 0.01 dBLevel accuracy 1 ±1.0 dBLevel linearity 1St<strong>and</strong>ard-20 <strong>to</strong> -15 dBm ±0.5 dB-15 <strong>to</strong> 0 dBm ±0.2 dB0 <strong>to</strong> +5 dBm ±0.5 dBOption 004-15 <strong>to</strong> +5 dBm ±0.2 dB+5 <strong>to</strong> +10 dBm ±0.5 dBImpedance50 Ω (nominal)<strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 3 <strong>GHz</strong> >18 dB RL (< 1.28 SWR)3 <strong>GHz</strong> <strong>to</strong> 6 <strong>GHz</strong> > 14 dB RL (


<strong>HP</strong> <strong>8753ET</strong> specificationsTest-p<strong>or</strong>t input characteristics (continued)External source mode 1 (CW time sweep only)Frequency range <strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 6 <strong>GHz</strong>R-channel input requirements 2Power level 0 <strong>to</strong> -25 dBmSpectral purityMaximum spurious


<strong>HP</strong> <strong>8753ET</strong>/ES supplementalcharacteristicsMeasurementNumber of display channelsFour display channels available.AccuracyThe following graph shows group-delay accuracy at1.3 <strong>GHz</strong> with type-N full two-p<strong>or</strong>t calibration <strong>and</strong>10-Hz IF b<strong>and</strong>width. Insertion loss is assumed <strong>to</strong> be< 2 dB <strong>and</strong> electrical length <strong>to</strong> be ten meters.Number of measurement channelsTwo primary <strong>and</strong> two auxiliary measurement channelsavailable.Measurement parameters<strong>HP</strong> <strong>8753ET</strong>: Reflection, transmission, A, B, R, A/R, B/R,A/B. Conversion <strong>to</strong> impedance <strong>or</strong> admittance.<strong>HP</strong> <strong>8753ES</strong>: S11, S21, S12, S22, A, B, R, A/R, B/R, A/B.Conversion <strong>to</strong> impedance <strong>or</strong> admittance.F<strong>or</strong>matsCartesian: log/linear magnitude, phase, group delay,SWR, real <strong>and</strong> imaginary.Smith chart: with log/linear amplitude <strong>and</strong> phase,R + jX, G + jB, <strong>or</strong> real/imaginary markers.Polar: with linear/log amplitude, phase, <strong>or</strong> real <strong>and</strong>imaginary markers.Data markersEach display channel has five independent markersthat can be displayed simultaneously. Twenty independentmarkers can be displayed in 4-channel displaymode when markers are uncoupled.Marker functionsMarkers can be used in various functions: Markersearch (Mkr <strong>to</strong> max, Mkr <strong>to</strong> min, Mkr <strong>to</strong> target), Mkrb<strong>and</strong>width with user-defined target values, Mkr <strong>to</strong>start, Mkr <strong>to</strong> s<strong>to</strong>p, Mkr <strong>to</strong> center, Mkr <strong>to</strong> span, Mkr <strong>to</strong>reference, Mkr <strong>to</strong> delay, <strong>and</strong> trace statistics (averagevalue, st<strong>and</strong>ard deviation, <strong>and</strong> peak-<strong>to</strong>-peak deviationof the data trace between two markers). The trackingfunction enables continuous update of marker searchvalues on each sweep.Group delay characteristicsAperture: selectableMaximum aperture: 20% of frequency spanMinimum aperture: (freq. span) / (number of points+1)RangeThe maximum delay is limited <strong>to</strong> measuring no m<strong>or</strong>ethan 180˚ of phase change within the minimum aperture.Range = 1 / (2 x minimum aperture)Source controlSweep limitsSet start/s<strong>to</strong>p <strong>or</strong> center/span of the stimulus parameter(frequency, power, time) directly through the sourcecontrol keys <strong>and</strong> the control knob, the step keys <strong>or</strong> thedata entry keyboard.Sweep typeSet a linear <strong>or</strong> logarithmic sweep, an arbitrarily definedfrequency list, a power sweep <strong>or</strong> a CW (single frequency)type of sweep.Measured number of points per sweepLinear frequency: choose 3, 11, 26, 51, 101, 201, 401,801, <strong>or</strong> 1601 points.Fast swept listDefine up <strong>to</strong> <strong>30</strong> different sub-sweep frequency rangesin any combination of CW, center/span, <strong>or</strong> start-s<strong>to</strong>psweep modes. Set test-p<strong>or</strong>t power levels <strong>and</strong> IF b<strong>and</strong>widthindependently f<strong>or</strong> each segment.Sweep modesSet a coupled channel sweep (same stimulus conditionson both channels) <strong>or</strong> an uncoupled channelsweep.Chop/alternateSelect whether <strong>to</strong> alternately <strong>or</strong> simultaneously (chop)measure channels when in dual-channel mode. Chopmode is faster, while alternate mode optimizes dynamicrange.Sweep timeSet sweep time in seconds, minutes <strong>or</strong> hours.Au<strong>to</strong>matic sweep timeSelect au<strong>to</strong> sweep time by entering zero secondssweep time. The analyzer will sweep at the minimumsweep time f<strong>or</strong> any subsequently selected stimulusconditions. Au<strong>to</strong> sweep time is the default condition.16


<strong>HP</strong> <strong>8753ET</strong>/ES supplementalcharacteristics (continued)Sweep triggerSet <strong>to</strong> either continuous, single, group sweep, <strong>or</strong> externaltrigger. Set external trigger <strong>to</strong> take a completesweep <strong>or</strong> <strong>to</strong> measure individual points in a frequency,power <strong>or</strong> list sweep.PowerSet source power from -20 <strong>to</strong> +5 dBm f<strong>or</strong> the<strong>HP</strong> <strong>8753ET</strong> 1 <strong>or</strong> from -85 <strong>to</strong> +10 dBm f<strong>or</strong> the<strong>HP</strong> <strong>8753ES</strong> 2 . Power slope can be set in dBm/<strong>GHz</strong>.Power meter calibrationSelect continuous leveling <strong>or</strong> use a c<strong>or</strong>rection table <strong>to</strong>modify source power. The c<strong>or</strong>rection table is createdwith an initial single sweep. Make single <strong>or</strong> multiplepower meter readings at each frequency.Data accuracy enhancementMeasurement calibrationMeasurement calibration significantly reduces measurementuncertainty due <strong>to</strong> err<strong>or</strong>s caused by systemdirectivity, source <strong>and</strong> load match, tracking, <strong>and</strong>crosstalk. Full two-p<strong>or</strong>t calibration removes all the systematicerr<strong>or</strong>s <strong>to</strong> obtain the most accurate measurements.Calibration types available• Frequency responseSimultaneous magnitude <strong>and</strong> phase c<strong>or</strong>rection of frequencyresponse err<strong>or</strong>s f<strong>or</strong> either reflection <strong>or</strong> transmissionmeasurements.• Response <strong>and</strong> isolationCompensates f<strong>or</strong> frequency response <strong>and</strong> directivity(reflection) <strong>or</strong> frequency response <strong>and</strong> crosstalkerr<strong>or</strong>s.• Enhanced response calibrationC<strong>or</strong>rects f<strong>or</strong> frequency response <strong>and</strong> source match f<strong>or</strong>transmission measurements, <strong>and</strong> provides one-p<strong>or</strong>tcalibration f<strong>or</strong> reflection measurements.• One-p<strong>or</strong>t calibrationUses test set p<strong>or</strong>t 1 <strong>or</strong> p<strong>or</strong>t 2 <strong>to</strong> c<strong>or</strong>rect f<strong>or</strong> directivity,frequency response <strong>and</strong> source match err<strong>or</strong>s.• Two-p<strong>or</strong>t calibrationCompensates f<strong>or</strong> directivity, source match, reflectionfrequency response, load match, transmission frequencyresponse <strong>and</strong> crosstalk f<strong>or</strong> an S-parameter test set.Crosstalk calibration can be omitted. Available on<strong>HP</strong> <strong>8753ES</strong> analyzers.• TRL*/LRM* calibrationCompensates f<strong>or</strong> directivity, reflection <strong>and</strong> transmissionfrequency response, <strong>and</strong> crosstalk in both the f<strong>or</strong>ward<strong>and</strong> reverse directions. Especially suitable f<strong>or</strong>calibrating non-coaxial environments, such as in testfixtures. TRL*/LRM* is a special implementation ofTRL/LRM calibration, modified f<strong>or</strong> the three-samplerreceiver in the <strong>HP</strong> <strong>8753ES</strong>. Available on <strong>HP</strong> <strong>8753ES</strong>analyzers.Interpolated err<strong>or</strong> c<strong>or</strong>rectionWith any type of accuracy enhancement applied, interpolatedmode recalculates the err<strong>or</strong> coefficients whenthe test frequencies are changed. The number ofpoints can be increased <strong>or</strong> decreased <strong>and</strong> thestart/s<strong>to</strong>p frequencies can be changed, but the resultingfrequency span must be a subset of the <strong>or</strong>iginal calibrationfrequency span. System perf<strong>or</strong>mance is notspecified f<strong>or</strong> measurements with interpolated err<strong>or</strong>c<strong>or</strong>rection applied.Velocity fac<strong>to</strong>rEnters the velocity fac<strong>to</strong>r <strong>to</strong> calculate equivalent electricallength.Reference plane extensionRedefine the plane-of-measurement reference <strong>to</strong> otherthan p<strong>or</strong>t 1 <strong>or</strong> p<strong>or</strong>t 2 of the <strong>HP</strong> <strong>8753ET</strong> <strong>and</strong> <strong>8753ES</strong>.Select default calibration kitSelect from a list of st<strong>and</strong>ard calibration kits <strong>or</strong> choosea user-defined kit.Data averagingIF b<strong>and</strong>width:The IF b<strong>and</strong>width is selectable from 6 <strong>kHz</strong> <strong>to</strong> 10 HzWeighted sweep-<strong>to</strong>-sweep averaging:Averages vec<strong>to</strong>r data on each successive sweep.Trace smoothingComputes the moving average of adjacent data points.Smoothing aperture defines the trace width (numberof points) <strong>to</strong> be averaged, <strong>and</strong> ranges from 0.25% <strong>to</strong>20% of the trace width.Display controlDisplay f<strong>or</strong>matsSingle-channel, dual-channel overlay (both traces onone graticule), dual-channel split (each trace on separategraticules), three-channel split (each trace onseparate graticules), three-channel overlay (threetraces on one graticule), quad-channel overlay (fourtraces on one graticule), quad-channel split (eachtrace on separate graticules).1. -85 <strong>to</strong> +5 dBm with Option 004.2. +8 dBm maximum with Option 075 <strong>or</strong> 014.17


<strong>HP</strong> <strong>8753ET</strong>/ES supplementalcharacteristics (continued)Trace functionsDisplay dataDisplay current measurement data, mem<strong>or</strong>y data, <strong>or</strong>current measurement with measurement <strong>and</strong> mem<strong>or</strong>ydata simultaneously.Trace mathVec<strong>to</strong>r division <strong>or</strong> subtraction of current linear measurementvalues <strong>and</strong> mem<strong>or</strong>y data.Display annotationsStart/s<strong>to</strong>p, center/span, <strong>or</strong> CW frequency, source level,scale/div, reference level, marker data, softkey functions,warning <strong>and</strong> caution messages, trace identification,<strong>and</strong> pass/fail indication.Au<strong>to</strong>scaleAu<strong>to</strong>matically selects scale resolution <strong>and</strong> referencevalue <strong>to</strong> center the trace.Electrical delayOffset measured phase <strong>or</strong> group delay by a definedamount of electrical delay, in seconds.Frequency blankingBlank out all frequency inf<strong>or</strong>mation on the display.Requires an instrument preset <strong>to</strong> re-enable frequencyinf<strong>or</strong>mation on the display.TitleAdd cus<strong>to</strong>m titles (49 characters maximum) <strong>to</strong> the display.Adjust displayCus<strong>to</strong>mize the col<strong>or</strong> <strong>and</strong> brightness of the data traces,mem<strong>or</strong>y traces, reference lines, graticules, text, <strong>and</strong>warning messages. Default col<strong>or</strong>s can be recalled alongwith one set of user-defined display values.S<strong>to</strong>rageInstrument stateUp <strong>to</strong> 31 instrument states can be s<strong>to</strong>red internally <strong>or</strong>recalled via the SAVE/RECALL menu. Instrumentstates include all control settings, active limit lines,active list frequency tables, mem<strong>or</strong>y trace data, activecalibration coefficients, <strong>and</strong> cus<strong>to</strong>m display titles.S<strong>to</strong>rage is in nonvolatile mem<strong>or</strong>y.Test sequencesSix measurement sequences can be s<strong>to</strong>red <strong>or</strong> recalledvia the sequencing menu. Sequences may also berecalled from Preset menu. Sequence register 6 is par<strong>to</strong>f nonvolatile s<strong>to</strong>rage <strong>and</strong> is not erased during a powercycle. If sequence 6 is titled AUTO, it will be executedwhen power is turned on.Disk driveData, instrument states, user graphics, data plots <strong>and</strong>test sequences can be s<strong>to</strong>red on internal floppy disk inMS-DOS ® <strong>or</strong> Hewlett-Packard's st<strong>and</strong>ard LIF f<strong>or</strong>mats.Data hardcopyData plottingHardcopy plots are au<strong>to</strong>matically produced with <strong>HP</strong>-GL compatible digital plotters. The <strong>HP</strong> <strong>8753ET</strong>/ES providesCentronics, RS-232C, <strong>and</strong> GP-IB interfaces.Data listingsPrin<strong>to</strong>uts of instrument data are directly producedwith a printer such as the <strong>HP</strong> DeskJet <strong>or</strong> LaserJet.Select black & white <strong>or</strong> col<strong>or</strong> print. F<strong>or</strong> a list ofcompatible printers, consult our printer-compatibilityguide Web page at http://www.hp.com/go/pcgConfigure plotsConfigure plots completely from the netw<strong>or</strong>k analyzerby defining pen col<strong>or</strong> <strong>and</strong> line type f<strong>or</strong> data, text markers,graticules, <strong>and</strong> mem<strong>or</strong>y traces.FunctionsPlot trace(s), graticule(s), marker(s), <strong>or</strong> text includingoperating <strong>and</strong> system parameters.QuadrantsPlot entire display in one of four different quadrants ofthe plotter paper.System capabilitiesLimit linesDefine test limit lines that appear on the display f<strong>or</strong>go/no go testing. Lines may be any combination of h<strong>or</strong>izontal,sloping lines, <strong>or</strong> discrete data points. Limit-testTTL output available f<strong>or</strong> external control <strong>or</strong> indication.External source modeThe receiver (input R) detects <strong>and</strong> phase-locks <strong>to</strong> anyexternally generated CW signal. Receiver inputs A <strong>and</strong>B will measure this same frequency f<strong>or</strong> comparison <strong>or</strong>tracking measurements.Au<strong>to</strong>maticThe input signal frequency is counted <strong>and</strong>displayed.ManualMeasures the input signal closest <strong>to</strong> the frequencyspecified by the user (within +0.5 <strong>to</strong> +5 MHz).Tuned receiverTunes the receiver f<strong>or</strong> a synthesized CW input signal ata precisely specified frequency. The time bases of theexternal RF source <strong>or</strong> sources must be tied <strong>to</strong> theexternal reference input (rear panel BNC). The builtinRF source is not used.18


<strong>HP</strong> <strong>8753ET</strong>/ES supplementalcharacteristics (continued)Frequency offset on/offSets the RF source <strong>to</strong> be swept at a fixed offset frequencyabove the receiver as required in a sweptRF/IF, fixed LO, mixer test.Service menuSelect the desired service test, service diagnostic, service<strong>or</strong> verification mode.Test sequencesDescriptionCreate, edit, save <strong>or</strong> recall a series of front-panel keystrokes<strong>to</strong> au<strong>to</strong>mate a measurement. Test sequencesmay contain basic stimulus <strong>and</strong> measurement functions(frequency, power, parameter, f<strong>or</strong>mat, scale)advanced operations (time domain, limit testing, displaymarker values) <strong>and</strong> basic logical branching (IFlimit test fails DO sequence 5 <strong>or</strong> GOSUB).S<strong>to</strong>rageTest sequences can be s<strong>to</strong>red internally <strong>to</strong> a disk drive<strong>and</strong> can be loaded from a computer over the GP-IBinterface. Sequence 6 is saved in nonvolatile s<strong>to</strong>rage<strong>and</strong> can be used as an au<strong>to</strong>start routine when titledAUTO.General purpose input/outputRead <strong>or</strong> write bits <strong>to</strong> the output p<strong>or</strong>t <strong>to</strong> control externaldevices such as part h<strong>and</strong>lers. Eight output <strong>and</strong>five input TTL lines are available on the parallel p<strong>or</strong>t.Other functionsPAUSE/continue, wait, title sequence, print sequence,duplicate sequence, pause <strong>and</strong> select.Time-domain (Option 010)With the time-domain option, data from transmission<strong>or</strong> reflection measurements in the frequency domainare converted <strong>to</strong> the time domain using a Fouriertransf<strong>or</strong>mation technique (chirp Z) <strong>and</strong> presented onthe display. The time-domain response shows the measuredparameter value versus time. Markers may alsobe displayed in electrical length (<strong>or</strong> physical length ifthe relative propagation velocity is entered).Time stimulus modesTwo types of time excitation stimulus wavef<strong>or</strong>ms canbe simulated during the transf<strong>or</strong>mations, a step <strong>and</strong> animpulse.• Low-pass stepThis stimulus, similar <strong>to</strong> a traditional time-domain reflec<strong>to</strong>meter(TDR) stimulus wavef<strong>or</strong>m, is used <strong>to</strong> measurelow-pass devices. The frequency-domain data shouldextend from DC (extrapolated value) <strong>to</strong> a higher value.• Low-pass impulseThis stimulus is also used <strong>to</strong> measure low-pass devices.• B<strong>and</strong>pass impulseThe b<strong>and</strong>pass impulse stimulates a pulsed RF signal(with an impulse envelope) <strong>and</strong> is used <strong>to</strong> measure thetime-domain response of b<strong>and</strong>-limited devices.WindowsThe windowing function can be used <strong>to</strong> modify (filter)the frequency-domain data <strong>and</strong> thereby reduce overshoot<strong>and</strong> ringing in the time-domain response. Threetypes of windows are available: minimum, n<strong>or</strong>mal, <strong>and</strong>maximum.GatingThe gating function can be used <strong>to</strong> selectively removereflection <strong>or</strong> transmission time-domain responses. Inconverting back <strong>to</strong> the frequency-domain the effects ofthe responses outside the gate are removed.Remote programmingInterface<strong>HP</strong>-IB interface operates <strong>to</strong> IEEE 488-1978 <strong>and</strong> IEC625 st<strong>and</strong>ards <strong>and</strong> IEEE 728-1982 recommended practices.AddressingThe GP-IB address can be verified <strong>or</strong> set from thefront panel via the local menu <strong>and</strong> can range from 0 <strong>to</strong><strong>30</strong> decimal (fac<strong>to</strong>ry set at 16).Pass controlAllows the <strong>HP</strong> <strong>8753ET</strong>/ES <strong>to</strong> request control of theGP-IB (when an active controller is present) wheneverit needs <strong>to</strong> output <strong>to</strong> a plotter <strong>or</strong> printer.System controllerLets an <strong>HP</strong> <strong>8753ET</strong>/ES become a controller on theGP-IB <strong>to</strong> directly control a plotter <strong>or</strong> a printer.Talker/listenerLets the <strong>HP</strong> <strong>8753ET</strong>/ES become an GP-IB talker/listenerwhen an external controller is present.Transfer f<strong>or</strong>matsBinary (internal 48-bit floating-point complex f<strong>or</strong>mat)ASCII 32- <strong>or</strong> 64-bit IEEE 754 floating-point f<strong>or</strong>matUser-accessible graphicsUsing a subset of <strong>HP</strong> Graphics Language (<strong>HP</strong>-GL),vec<strong>to</strong>r <strong>or</strong> text graphics may be written on the<strong>HP</strong> <strong>8753ET</strong>/ES via GP-IB. Up <strong>to</strong> 5 kbytes of data canbe s<strong>to</strong>red at one time (4 bytes per vec<strong>to</strong>r, 2 bytes percharacter).Interface function codesSH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP0, DC1,DT0, C1, C2, C3, C10, E219


General characteristicsFront panel connec<strong>to</strong>rs<strong>HP</strong> <strong>8753ES</strong> test p<strong>or</strong>ts (without Option 011)Connec<strong>to</strong>r type 7 mm, precisionImpedance 50 ohms (nominal)<strong>HP</strong> <strong>8753ES</strong> Option 011 test p<strong>or</strong>tsConnec<strong>to</strong>r type Type-N femaleImpedance 50 ohms (nominal)<strong>HP</strong> <strong>8753ES</strong> Option 075 test p<strong>or</strong>tsConnec<strong>to</strong>r type Type-N femaleImpedance 75 ohms (nominal)<strong>HP</strong> <strong>8753ET</strong> test p<strong>or</strong>tsConnec<strong>to</strong>r typeImpedanceProbe powerRear-panel connec<strong>to</strong>rsType-N female50 ohms (nominal)+15V ±2% 400 mA(combined load f<strong>or</strong> bothprobe connections)-12.6V ±5.5% <strong>30</strong>0 mA(combined load f<strong>or</strong> bothprobe connections)External reference frequency input (EXT REF INPUT)Frequency 1, 2, 5, <strong>and</strong> 10 MHz(±200 Hz at 10 MHz)Level-10 dBm <strong>to</strong> +20 dBm, (typical)Impedance 50 ohmsConnec<strong>to</strong>r BNC (f)High-stability frequency reference output(Option 1D5)Frequency 10.0000 MHzFrequency stability ±0.05 ppm(0 ˚C <strong>to</strong> 55 ˚C)Daily aging rate(after <strong>30</strong> days)


Environmental characteristicsGeneral conditionsRFI <strong>and</strong> EMI susceptibility: defined by VDE 07<strong>30</strong>,CISPR Publication 11, <strong>and</strong> FCC Class B St<strong>and</strong>ards.ESD (electrostatic discharge): must be eliminated byuse of static-safe w<strong>or</strong>k procedures <strong>and</strong> an anti-staticbench mat. The flexible rubber keypad protects keycontacts from dust, but the environment should be asdust-free as possible f<strong>or</strong> optimal reliability.Operating conditionsTemperature(unless otherwise noted) 0˚ <strong>to</strong> 55 ˚CHumidity5% <strong>to</strong> 95% at 40 ˚C(non-condensing)Altitude0 <strong>to</strong> 4500 meters(15,000 feet)Non-operating s<strong>to</strong>rage conditionsTemperatureHumidityAltitudeWeight (<strong>HP</strong> <strong>8753ET</strong>/ES)NetShipping-40 ˚C <strong>to</strong> +70 ˚C0 <strong>to</strong> 90% relative at+65 ˚C (noncondensing)0 <strong>to</strong> 15,240 meters(50,000 feet)21 kg (46 lb)35 kg (77 lb)Cabinet dimensions (<strong>HP</strong> <strong>8753ET</strong>/ES)(These dimensions exclude front <strong>and</strong> rear panelprotrusions.)222 mm H x 425 mm W x 457 mm D(8.75 in x 16.75 in x 18.0 in)Measurement throughput summaryThe following table shows typical measurement timesin milliseconds.Typical time f<strong>or</strong> completion (msec)Number of PointsMeasurement 51 201 401 1601Start=1 <strong>GHz</strong>, Span=10 MHz, IFBW=6 <strong>kHz</strong>Unc<strong>or</strong>rected, 1-p<strong>or</strong>t calibration: 1 32 70 121 423Two-p<strong>or</strong>t calibration: 2 62 139 240 848Start=<strong>30</strong> <strong>kHz</strong>, S<strong>to</strong>p=3 <strong>GHz</strong>, IFBW=6 <strong>kHz</strong>Unc<strong>or</strong>rected, 1-p<strong>or</strong>t calibration 1 202 270 <strong>30</strong>4 615Two-p<strong>or</strong>t calibration 2 402 540 607 1237Start=<strong>30</strong> <strong>kHz</strong>, S<strong>to</strong>p=6 <strong>GHz</strong>, IFBW=6 <strong>kHz</strong>Unc<strong>or</strong>rected, 1-p<strong>or</strong>t calibration 1 310 380 415 658Two-p<strong>or</strong>t calibration 2 618 757 829 1315Time domain (increase over unc<strong>or</strong>rected sweep time) 3Conversion 12 42 86 378Gating 14 40 80 349GP-IB data transfer 4Internal binary 10 16 21 58ASCII 35 112 214 831IEEE 754 floating point f<strong>or</strong>mat32-bit 11 19 28 8364-bit 13 26 42 141OptionsHarmonic measurements (Option 002)Measures amplifier 2nd <strong>and</strong> 3rd harmonics on a sweptfrequencybasis f<strong>or</strong> fundamental signals above 16 MHz.Dynamic range (source at +10 dBm,receiver


Test set specifications f<strong>or</strong> Option 011<strong>HP</strong> 85046A/B S-parameter test setsThe <strong>HP</strong> 85046A/B S-parameter test sets provide thecapability <strong>to</strong> measure reflection <strong>and</strong> transmission characteristics(including S-parameters) of two-p<strong>or</strong>tdevices in either direction with a single connection.The test sets are controlled from the <strong>HP</strong> <strong>8753ES</strong>Option 011 <strong>and</strong> include a programmable step attenua<strong>to</strong>r.The frequency range of the <strong>HP</strong> 85046A 50-ohmtest set is <strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 3 <strong>GHz</strong>. The <strong>HP</strong> 85046A has precision7-mm connec<strong>to</strong>rs. The frequency range of the<strong>HP</strong> 85046B 75-ohm test set is <strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 2 <strong>GHz</strong>. The<strong>HP</strong> 85046B has 75-ohm type-N(f) connec<strong>to</strong>rs. Bothconnec<strong>to</strong>rs can be adapted <strong>to</strong> other interfaces withthe appropriate precision adapters.<strong>HP</strong> 85046A/B SpecificationsNote: Specifications that apply only <strong>to</strong> the <strong>HP</strong> 85046are indicated in parentheses.Impedance50 ohm (75 ohm)Frequency range <strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 3 <strong>GHz</strong>Directivity(<strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 2 <strong>GHz</strong>)35 dB <strong>to</strong> 1.3 <strong>GHz</strong><strong>30</strong> dB <strong>to</strong> F max1Typical trackingTransmission magnitude, phase 20.3 MHz <strong>to</strong> 2.0 MHz ±1.5 dB, ±20˚2.0 MHz <strong>to</strong> F max ±1.5 dB, ±10˚Reflection magnitude, phase 20.3 MHz <strong>to</strong> 2.0 MHz ±1.5 dB, ±25˚2.0 MHz <strong>to</strong> F max ±1.5 dB, ±10˚Effective source match0.3 MHz <strong>to</strong> 2.0 MHz 14 dB2.0 MHz <strong>to</strong> 1.3 <strong>GHz</strong> 20 dB (17 dB)1.3 <strong>GHz</strong> <strong>to</strong> F max 16 dBNominal insertion lossInput <strong>to</strong> test p<strong>or</strong>t14 dB + 0.5 dB/<strong>GHz</strong>(19.5 dB + 1 dB/<strong>GHz</strong>)Input <strong>to</strong> incident 18 dB + 1.5 dB/<strong>GHz</strong>(18 dB + 1.5 dB/<strong>GHz</strong>)P<strong>or</strong>t 1, 2 <strong>to</strong> A, B 6.5 dB + 1.0 dB/<strong>GHz</strong>(12 dB + 0.5 dB <strong>GHz</strong>)Test set switch/repeatability±0.03 dB; ±0.01 dB 3Max. operating level +20 dBmDamage level +<strong>30</strong> dBmRF attenua<strong>to</strong>r range 70 dB, 10 dB stepsDC bias range ±<strong>30</strong> VDC, 200 mA 4500 mA maxA st<strong>and</strong>ard <strong>HP</strong> 85046A/B test set contains a solid-statetransfer switch, which allows continuous switching ofpower from p<strong>or</strong>t 1 <strong>to</strong> p<strong>or</strong>t 2 f<strong>or</strong> full two-p<strong>or</strong>t err<strong>or</strong> c<strong>or</strong>rection.Option 009 replaces the transfer switch with amechanical switch. This provides about 1.5 dB m<strong>or</strong>epower at the test p<strong>or</strong>t, but does not allow continuousswitching, so the user must initiate updates of all fourS-parameters f<strong>or</strong> full two-p<strong>or</strong>t err<strong>or</strong> c<strong>or</strong>rection. Also,the mechanical switch has relays that will wear outfaster than the solid-state switch. Approximate lifetimeof the mechanical switch is 1 million cycles.<strong>HP</strong> 85046A schematicDC bias connec<strong>to</strong>rsIncludesDimensionsWeight50 ohm BNC (f)four 190 mm (7.5 in) type-Ncables <strong>and</strong> test setinterconnect cable.90 mm H x 432 mm W x 553 mm D9.1 kg (20 lb)1. F max is the upper frequency limit of the associated test set.2. Degrees, specified as deviation from linear phase.3. Typical perf<strong>or</strong>mance.4. Some degradation of RF specifications may occur.22


<strong>HP</strong> 85047A S-parameter test setThe <strong>HP</strong> 85047A S-parametercapability <strong>to</strong> simultaneously<strong>and</strong> transmissioneither direction withincludes a frequencyan <strong>HP</strong> 8753B/C<strong>GHz</strong> in a single sweep<strong>kHz</strong> <strong>to</strong> 3 <strong>GHz</strong> in aOption 011 does<strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 6exhibits


Access<strong>or</strong>iesCalibration kitsVec<strong>to</strong>r accuracy enhancement procedures require thatthe systematic err<strong>or</strong>s of the measurement system becharacterized by measuring known devices (st<strong>and</strong>ards)on the system over the frequency range of interest.Return loss specifications <strong>or</strong> typical values are providedwhere available f<strong>or</strong> the terminations <strong>and</strong> adapters.<strong>HP</strong> 85031B 7-mm calibration kitContains precision 7-mm st<strong>and</strong>ards used <strong>to</strong> calibratethe <strong>HP</strong> <strong>8753ES</strong> f<strong>or</strong> measurement of devices with precision7-mm connec<strong>to</strong>rs.Includes<strong>HP</strong> part number7 mm sh<strong>or</strong>t/open circuit 85031-600017 mm 50-ohm terminations(two each) 00909-60008Specifications f<strong>or</strong> terminationsDC <strong>to</strong> 5 <strong>GHz</strong> RL ≥ 52 dB5 <strong>to</strong> 6 <strong>GHz</strong> RL ≥ 46 dB<strong>HP</strong> 85032B 50-ohm type-N calibration kitContains precision 50-ohm type-N st<strong>and</strong>ards used <strong>to</strong>calibrate the <strong>HP</strong> <strong>8753ET</strong>/ES <strong>and</strong> 50-ohm test sets f<strong>or</strong>measurement of devices with 50-ohm type-N connec<strong>to</strong>rs.Precision phase-matched 7-mm <strong>to</strong> type-Nadapters are included f<strong>or</strong> accurate measurements ofnon-insertable devices. Option 001 deletes the 7-mm<strong>to</strong> type-N adapters, which are not needed with the<strong>HP</strong> <strong>8753ET</strong>.Includes<strong>HP</strong> part numberN-male 50-ohm termination 00909-60009N-female 50-ohm termination 00909-60010N-male sh<strong>or</strong>t circuit 85032-60008N-female sh<strong>or</strong>t circuit 85032-60009N-female open circuit 85032-60012N-male open circuit 85032-600077-mm <strong>to</strong> N-male adapter(two each) 85054-600097-mm <strong>to</strong> N-female adapter(two each) 85054-60001Specifications f<strong>or</strong> terminations:DC <strong>to</strong> 3 <strong>GHz</strong> RL ≥ 49 dB2 <strong>to</strong> 3 <strong>GHz</strong> RL ≥ 46 dB3 <strong>to</strong> 6 <strong>GHz</strong> RL ≥ 40 dBTypical adapter characteristics:DC <strong>to</strong> 6 <strong>GHz</strong> RL ≥ <strong>30</strong> dB<strong>HP</strong> 85033D 3.5-mm calibration kitContains a set of precision 3.5-mm st<strong>and</strong>ards <strong>to</strong> calibratethe <strong>HP</strong> <strong>8753ES</strong> <strong>and</strong> 50-ohm test sets f<strong>or</strong> themeasurement of devices with precision 3.5-mm <strong>and</strong>SMA connec<strong>to</strong>rs. Precision phase-matched 7-mm <strong>to</strong>3.5-mm adapters are included f<strong>or</strong> accurate measurementsof non-insertable devices. Option 001 deletesthe 7-mm <strong>to</strong> type-N adapters.Includes<strong>HP</strong> part number3.5-mm-male 50-ohm termination 85033-600093.5-mm-female 50-ohm termination 85033-600103.5-mm-female sh<strong>or</strong>t 85033-600143.5-mm-male sh<strong>or</strong>t 85033-600133.5-mm-female open 85033-600123.5-mm-male open 85033-600117-mm <strong>to</strong> 3.5-mm femaleadapter (two) 1250-17477-mm <strong>to</strong> 3.5-mm maleadapter (two) 1250-1746Specifications f<strong>or</strong> terminations:DC <strong>to</strong> 1.3 <strong>GHz</strong>RL ≥ 46 dB1.3 <strong>to</strong> 3 <strong>GHz</strong> RL ≥ 44 dB3 <strong>to</strong> 6 <strong>GHz</strong> RL ≥ 38 dBTypical adapter characteristics:DC <strong>to</strong> 6 <strong>GHz</strong>RL ≥ 34 dB<strong>HP</strong> 85036B 75-ohm type-N calibration kitContains a set of precision 75-ohm type-N st<strong>and</strong>ards <strong>to</strong>calibrate the <strong>HP</strong> <strong>8753ES</strong> with a 75-ohm test set f<strong>or</strong>measurement of devices with 75-ohm type-N connec<strong>to</strong>rs.Precision phased-matched adapters are includedf<strong>or</strong> accurate measurements of non-insertable devices.Includes<strong>HP</strong> part numberN-male 75-ohm termination 00909-60019N-female 75-ohm termination 00909-60020N-female 75-ohm sh<strong>or</strong>t 85036-60011N-male 75-ohm sh<strong>or</strong>t 85036-60012N-female open 85032-20001N-male open 85032-60007N-male <strong>to</strong> N-male 75-ohm adapter 85036-60013N-female <strong>to</strong> N-female 75-ohm adapter 85036-60014N-male <strong>to</strong> N-female 75-ohm adapter 85036-60015Specifications f<strong>or</strong> terminations:DC <strong>to</strong> 2 <strong>GHz</strong>RL ≥ 46 dB2 <strong>to</strong> 3 <strong>GHz</strong> RL ≥ 40 dB24


Verification kitsMeasuring known devices other than the st<strong>and</strong>ardsused in calibration is an easy way <strong>to</strong> verify the properoperation of an <strong>HP</strong> <strong>8753ES</strong> measurement system. <strong>HP</strong>offers verification kits which include devices, withdata, f<strong>or</strong> verifying the err<strong>or</strong>-c<strong>or</strong>rected measurementsof an <strong>HP</strong> <strong>8753ES</strong> <strong>and</strong> 50-ohm test sets.<strong>HP</strong> 85029B 7-mm verification kitContains a set of precision 7-mm devices, with datatraceable <strong>to</strong> NIST* f<strong>or</strong> verifying the calibrated perf<strong>or</strong>manceof an <strong>HP</strong> 8753E measurement system.The <strong>HP</strong> 85031B 7-mm calibration kit is required f<strong>or</strong>complete verification.Test-p<strong>or</strong>t return cablesHewlett-Packard offers high-quality RF cables <strong>to</strong> connectthe <strong>HP</strong> <strong>8753ET</strong>/ES <strong>and</strong> test sets <strong>to</strong> devices undertest. These cables offer excellent RF shielding f<strong>or</strong>high- dynamic-range measurements.<strong>HP</strong> 11851B 50-ohm type-N RF cable kitRecommended f<strong>or</strong> use with the <strong>HP</strong> 11850C/D threewaypower splitters. Kit includes three phase-matched610-mm (24-in.) cables <strong>and</strong> one 860-mm (34-in.)cable.Return loss> 24 dB <strong>to</strong> 3 <strong>GHz</strong>Phase tracking ±4˚ at 1.3 <strong>GHz</strong><strong>HP</strong> 8120-5639 50-ohm type-N RF cableA 50-ohm type-N test p<strong>or</strong>t cable (both connec<strong>to</strong>rs aretype-N male).Return loss> <strong>30</strong>dB <strong>to</strong> 6 <strong>GHz</strong><strong>HP</strong> 11857B 75-ohm type-N test p<strong>or</strong>treturn cablesA pair of 610-mm (24-in.) test p<strong>or</strong>t return cables f<strong>or</strong>use with the <strong>HP</strong> <strong>8753ES</strong> <strong>or</strong> <strong>HP</strong> 85046B 75-ohm S-parametertest set.Return lossPhase tracking> 24 dB <strong>to</strong> 2 <strong>GHz</strong>±2˚ at 1.3 <strong>GHz</strong><strong>HP</strong> 11857D 7-mm test-p<strong>or</strong>t return cablesA pair of 610-mm (24-in.) test p<strong>or</strong>t return cables f<strong>or</strong>use with the <strong>HP</strong> <strong>8753ES</strong> with the <strong>HP</strong> 85046A <strong>and</strong>85047A S-parameter test sets. These cables can beused with connec<strong>to</strong>r types other than 7-mm with theappropriate precision adapters.Return lossPhase tracking> 24 dB <strong>to</strong> 3 <strong>GHz</strong>> 20 dB <strong>to</strong> 6 <strong>GHz</strong>±2˚ at 1.3 <strong>GHz</strong><strong>HP</strong> 11857F type-F test-p<strong>or</strong>t return cablesType-F test p<strong>or</strong>t cable set (75-ohm). Includes onecable with type-N(male) <strong>to</strong> type-F(male) connec<strong>to</strong>rs<strong>and</strong> one cable with type-N(male) <strong>to</strong> type-F(female)connec<strong>to</strong>rs.<strong>HP</strong> 11850C/D three-way power splitters<strong>HP</strong> 11850C <strong>HP</strong> 11850DImpedance 50 ohms 75 ohmsFrequency range DC <strong>to</strong> 3 <strong>GHz</strong> DC <strong>to</strong> 2 <strong>GHz</strong>Tracking ±25 dB, ±3˚ ±2 dB, ±2.5˚Equivalentsource match <strong>30</strong> dB at 1.3 <strong>GHz</strong> <strong>30</strong> dB at 1.3 <strong>GHz</strong>(ratio <strong>or</strong> leveling) 20 dB at 3 <strong>GHz</strong> 20 dB at 2 <strong>GHz</strong>Nominal9.5 dB + 1 dB/<strong>GHz</strong> 7.8 dBinsertion lossInput p<strong>or</strong>t matchDC <strong>to</strong> 1.3 <strong>GHz</strong> 20 dB 20 dB1.3 <strong>GHz</strong> <strong>to</strong> F max 10 dB 10 dBMaximumoperating level +20 dB +20 dBDamage level +<strong>30</strong> dB +<strong>30</strong> dBRF connec<strong>to</strong>rsRF input 50-ohm type-N (f) 50-ohm type-N(f)All others 50-ohm type-N (f) 75-ohm type-N(f)IncludesThree <strong>HP</strong> 11852B 50-<strong>to</strong> 75-ohmminimum-loss padsRecommended access<strong>or</strong>ies<strong>HP</strong> 11851B RF cable kit<strong>HP</strong> 11667A 50-ohm power splitterFrequency range DC <strong>to</strong> 18 <strong>GHz</strong>Typical insertion loss 6 dBEquivalent source matchDC <strong>to</strong> 4 <strong>GHz</strong>26 dB4 <strong>GHz</strong> <strong>to</strong> 8 <strong>GHz</strong> 21 dB8 <strong>GHz</strong> <strong>to</strong> 18 <strong>GHz</strong> 17 dBTracking(between output arms)DC <strong>to</strong> 4 <strong>GHz</strong>±15 dB4 <strong>GHz</strong> <strong>to</strong> 8 <strong>GHz</strong> ±2 dB8 <strong>GHz</strong> <strong>to</strong> 18 <strong>GHz</strong> ±25 dBMaximum operating level ±27 dBmConnec<strong>to</strong>rs50-ohm type-N (f)Option 001RF inputtype-N (m)Outputstype-N (f)Option 002RF inputtype-N (f) on RFOutputsprecision 7-mmDimensions46 mm H x 52 mmW x19 mm D(1.8 x 2.0 x 0.7 in)Recommended access<strong>or</strong>ies<strong>HP</strong> 11851B RF cable kit<strong>HP</strong> 11852B 50-ohm <strong>to</strong> 75-ohm minimum loss padFrequency rangeNominal insertion lossReturn loss<strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 2 <strong>GHz</strong> 32 dBDC <strong>to</strong> 3.0 <strong>GHz</strong>5.7 dB2 <strong>GHz</strong> <strong>to</strong> 3 <strong>GHz</strong> 27 dBMaximum input power 250 mW (+24 dBm)26


<strong>HP</strong> 11852B 50-ohm <strong>to</strong> 75-ohm minimum loss pad(continued)Connec<strong>to</strong>rsDimensionsWeight50-ohm access<strong>or</strong>y kits50-ohm type-N (f) <strong>to</strong>75-ohm type-N (m)st<strong>and</strong>ard Option 004,50-ohm type-N (m) <strong>to</strong>75-ohm type-N (f)14-mm D x 70-mm L(0.56 in x 2.75 in)Net 0.1 kg (0.316 lb)The <strong>HP</strong> 11853A 50-ohm type-N <strong>and</strong> the <strong>HP</strong> 11854A50-ohm BNC access<strong>or</strong>y kits provide the RF componentsgenerally required when using either the<strong>HP</strong> 85046A, 85047A <strong>or</strong> 11850C with the <strong>HP</strong> <strong>8753ES</strong>Option 011 when measuring devices having 50-ohmtype-N <strong>or</strong> BNC connec<strong>to</strong>rs. These kits are suppliedwith a s<strong>to</strong>rage case.<strong>HP</strong> 11853A 50-ohm type-N access<strong>or</strong>y kitIncludes<strong>HP</strong> part numberType-N (f) sh<strong>or</strong>t<strong>HP</strong> 11511AType-N (m) sh<strong>or</strong>t<strong>HP</strong> 11512AType-N (m) <strong>to</strong> N (m) adapter 1250-1475Type-N (f) <strong>to</strong> N (f) adapter 1250-1472<strong>HP</strong> 11854A 50-ohm BNC access<strong>or</strong>y kitIncludes<strong>HP</strong> part numberType-N (m) <strong>to</strong> BNC female adapter 1250-1476Type-N (m) <strong>to</strong> BNC male adapter 1250-1473Type-N (f) <strong>to</strong> BNC male adapter 1250-1477Type-N (f) <strong>to</strong> BNC female adapter 1250-1474BNC (m) sh<strong>or</strong>t 1250-092975-ohm access<strong>or</strong>y kitsThe <strong>HP</strong> 11855A 75-ohm type-N <strong>and</strong> the <strong>HP</strong> 11856A75-ohm BNC access<strong>or</strong>y kits provide the RF componentsgenerally required when using either the<strong>HP</strong> 85046B <strong>or</strong> the <strong>HP</strong> 11850D power splitter with the<strong>HP</strong> <strong>8753ES</strong> Option 011 when measuring devices having75-ohm type-N <strong>or</strong> BNC connec<strong>to</strong>rs. These kits are suppliedwith a s<strong>to</strong>rage case.<strong>HP</strong> 11855A 75-ohm type-N access<strong>or</strong>y kitIncludes<strong>HP</strong> part numberType-N (f) sh<strong>or</strong>t 1250-1531Type-N (m) sh<strong>or</strong>t 1250-15<strong>30</strong>Type-N (m) <strong>to</strong> N (m) adapter 1250-1528Type-N (f) <strong>to</strong> N (f) adapter 1250-1529Type-N (m) termination 1250-1532<strong>HP</strong> 11856A 75-ohm BNC access<strong>or</strong>y kitIncludes<strong>HP</strong> part numberType-N (m) <strong>to</strong> BNC (f) adapter 1250-1535Type-N (m) <strong>to</strong> BNC (m) adapter 1250-1533Type-N (f) <strong>to</strong> BNC (m) adapter 1250-1534Type-N (f) <strong>to</strong> BNC (f) adapter 1250-1536BNC (m) sh<strong>or</strong>t 1250-0929BNC (m) termination 11652-600101. Return loss <strong>and</strong> insertion loss limited below 16 MHz by series capaci<strong>to</strong>r.2. Internal bias tees cannot be used with this limiter.RF limiterExternally attaches <strong>to</strong> one <strong>or</strong> both p<strong>or</strong>ts of the analyzer.Provides protection against potential high-powertransients from external devices.Specifications<strong>HP</strong> 119<strong>30</strong>A 7-mm RF limiterFrequency range DC <strong>to</strong> 6 <strong>GHz</strong>Nominal insertion lossDC <strong>to</strong> 3 <strong>GHz</strong> 1.0 dB3 <strong>GHz</strong> <strong>to</strong> 6 <strong>GHz</strong> 1.5 dBReturn lossDC <strong>to</strong> 3 <strong>GHz</strong> 22 dB3 <strong>GHz</strong> <strong>to</strong> 6 <strong>GHz</strong> 20 dBMaximum input power 3WMaximum DC<strong>30</strong> V, 350 mA<strong>HP</strong> 119<strong>30</strong>B 50-ohm type-N RF limiter 2Frequency range 5 MHz <strong>to</strong> 6 <strong>GHz</strong>Nominal insertion loss5 MHz <strong>to</strong> 3 <strong>GHz</strong> 1.0 dB 13 <strong>GHz</strong> <strong>to</strong> 6 <strong>GHz</strong> 1.5 dBReturn loss5 MHz <strong>to</strong> 3 <strong>GHz</strong> 21 dB 13 <strong>GHz</strong> <strong>to</strong> 6 <strong>GHz</strong> 17 dBMaximum input power 3W<strong>HP</strong> 85024A high-frequency probeTwo probes may be powered directly from the front panelof the <strong>HP</strong> <strong>8753ET</strong>/ES. Refer <strong>to</strong> technical specifications,<strong>HP</strong> literature #5954-8393, f<strong>or</strong> m<strong>or</strong>e inf<strong>or</strong>mation.SpecificationsInput capacitance (at 500 MHz)


<strong>HP</strong> 8347A RF amplifierThe <strong>HP</strong> 8347A RF amplifier delivers increased power across a<strong>30</strong>0 <strong>kHz</strong> <strong>to</strong> 3 <strong>GHz</strong> frequency range.The <strong>HP</strong> 8347A provides leveled output power without using anexternal coupler <strong>and</strong> detec<strong>to</strong>r, since these parts are built-in. Theexternal ALC can be directly connected <strong>to</strong> the external AM input onthe <strong>HP</strong> <strong>8753ET</strong>/ES. This capability is especially useful f<strong>or</strong> achievinghigh dynamic range measurements at faster sweep rates.SpecificationsFrequency100 <strong>kHz</strong> <strong>to</strong> 3 <strong>GHz</strong>Gain25 dB minimumOutput power (leveled) +5 dBm <strong>to</strong> +20 dBm(adjustable)Maximum output power 24 dBm (typical)Leveled power flatness ±1.5 dBImpedance50 ohms nominalSWRInput2.2:1 maxOutput1.6:1 (ALC on)Spectral purityHarmonics-20 dBc at dBmThird-<strong>or</strong>der intercept +<strong>30</strong> dBm (nominal)Typical noise figure100 MHz <strong>to</strong> 3 <strong>GHz</strong> 13.5 dBRF connec<strong>to</strong>rsType-N femaleDimensions102-mm H x 213-mmW x 297-mm D(4.0 in x 8.4 in x 11.7 in)Weightnet 3.5 kg (7.7 lb)F<strong>or</strong> m<strong>or</strong>e inf<strong>or</strong>mation aboutHewlett-Packard test <strong>and</strong> measurementproducts, applications,services, <strong>and</strong> f<strong>or</strong> a current salesoffice listing, visit our web site,http://www.hp.com/go/tmdir. Youcan also contact one of the followingcenters <strong>and</strong> ask f<strong>or</strong> a test <strong>and</strong>measurement sales representative.United States:Hewlett-Packard CompanyTest <strong>and</strong> Measurement Call CenterP.O. Box 4026Englewood, CO 80155-40261 800 452 4844Canada:Hewlett-Packard Canada Ltd.5150 Spectrum WayMississauga, Ontario L4W 5G1(905) 206 4725Europe:Hewlett-PackardEuropean Marketing CentreP.O. Box 9991180 AZ AmstelveenThe Netherl<strong>and</strong>s(31 20) 547 9900Japan:Hewlett-Packard Japan Ltd.Measurement Assistance Center9-1, Takakura-Cho, Hachioji-Shi,Tokyo 192, JapanTel: (81) 426-56-7832Fax: (81) 426-56-7840Latin America:Hewlett-PackardLatin American Region Headquarters5200 Blue Lagoon Drive, 9th Flo<strong>or</strong>Miami, Fl<strong>or</strong>ida 33126, U.S.A.(<strong>30</strong>5) 267 4245/4220Australia/New Zeal<strong>and</strong>:Hewlett-Packard Australia Ltd.31-41 Joseph StreetBlackburn, Vic<strong>to</strong>ria 31<strong>30</strong>, Australia1 800 629 485Asia Pacific:Hewlett-Packard Asia Pacific Ltd19/F Cityplaza One1111 King's RoadTaikoo Shing, Hong Kongtel: 852-2599-7777fax: 852-2506-9285Data Subject <strong>to</strong> ChangeCopyright © 1999Hewlett-Packard CompanyPrinted in U.S.A. 8/995968-5160EMS-DOS is a U.S. registered trademark of Microsoft C<strong>or</strong>p<strong>or</strong>ation.Pentium is a U.S. registered trademark of Microsoft C<strong>or</strong>p<strong>or</strong>ation.28

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!