84 Pharmacopeia vs. Thermodynamic <strong>Melting</strong> <strong>Point</strong>s Appendix Aresults, and within the accuracy of the measurement (±0.3 ºC), the thermodynamicmelting point results are independent of the ramp rate as expected.Ramp Rate, r[°C/min]Clear <strong>Point</strong> [°C](MP pharma )MP therm = MP pharma –ThermoCF . r 1/2[°C]0.1 134.2 133.60.2 134.4 133.40.5 134.9 133.61 135.4 133.52 136.2 133.55 137.9 133.6Table 7. Clear and Thermodynamic <strong>Melting</strong> <strong>Point</strong> of Phenacetin at various ramp rates(ThermoCF = 1.9).TipThermodynamic corrections, ∆T(r), are automatically appended to melting point Reportsif Thermo MP is set to Calculate in the Options menu of the Melt Screen.<strong>OptiMelt</strong> <strong>Automated</strong> <strong>Melting</strong> <strong>Point</strong> <strong>System</strong>