Chapter 4 Maintenance and Calibration 65Does Your <strong>OptiMelt</strong>Need Calibration?Is the current calibration trusted?NoYesIs the current calibration GLPdocumented?NoYesHas the current calibration expired?YesNoHas the heating block beenmodified since the last calibration?YesNoIs a new ramping rate required?YesNoYour unit is Acceptable formelting point determinationsPerform Acceptability Test andcalculate temperature offsetcorrections (TOCs)YesDo the TOCs meet specifications?NoUpdate the unit’s temperatureoffsets using the calculated TOCsFigure 35. Temperature Offset Calibration Protocol.<strong>OptiMelt</strong> <strong>Automated</strong> <strong>Melting</strong> <strong>Point</strong> <strong>System</strong>