50 <strong>Melting</strong> <strong>Point</strong> Determination Chapter 3Instrument SetupAlong with proper sample preparation, careful selection of the instrumentsettings is also essential for accurate and reproducible melting pointdeterminations.The modern trend in melting point instrumentation is towards small aluminum ovens. Atypical oven can hold three capillaries and the thermal mass around the three tubes is veryclose so that deviations as small as 0.02 to 0.1ºC (temperature dependent) are keptbetween the three tubes during a melt.The main advantage of a small metal oven is the lack of an overshoot that lets you parkthe unit at a start temperature