16 Getting Started Chapter 1• Touch the Graph button on the Report screen to display the Melt Graphscorresponding to the Report being displayed.• Melt Graphs are included in all hardcopy printouts of Reports generated with printerssupporting graphical output.• Real time Melt Graphs are also available through the MeltView software datadisplay.The camera threshold values (Onset%, Single% and Clear%) used by the Digital ImageProcessor to determine melting temperatures are user-programmable. The factory defaultvalues loaded into your <strong>OptiMelt</strong> are generally adequate for most tests and samples,however, fine tuning of their values might be required to better match your visual andautomated results. Follow the instructions in Chapter 4 to view and modify the camerathresholds.<strong>OptiMelt</strong> <strong>Automated</strong> <strong>Melting</strong> <strong>Point</strong> <strong>System</strong>