System Introduction Applied Biosystems 7300 &7500 Real ... - CRN2M
System Introduction Applied Biosystems 7300 &7500 Real ... - CRN2M
System Introduction Applied Biosystems 7300 &7500 Real ... - CRN2M
- No tags were found...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Monthly MaintenanceIf the pixel intensity shows a bright spot in the ROI Inspectorview, and background raw data exceeds 72.000 rfu, check thethermal cycler block for contamination:• Leave the block empty• Open ROI Inspector• Block up• Take Snapshots43¢¡¤£ £ ¥§¦©¨ ¨ ROI Inspector• Open a new document• Instrument > Calibrate2048 msBlock upFilter ASnapshot44If any well shows a bright spot, run abackground run (see following slides)¢¡¤£ £ ¥§¦©¨ ¨