11.07.2015 Views

ACE C18-AR and C18-PFP - Hplc.eu

ACE C18-AR and C18-PFP - Hplc.eu

ACE C18-AR and C18-PFP - Hplc.eu

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

®<strong>ACE</strong>HPLC / UHPLC ColumnsSummary: Unwanted Selectivity Changes44Pressure induced k <strong>and</strong> changes may be seen forany manufacturer phases under UHPLC conditionsChanges in selectivity <strong>and</strong> retention may be significantwith ionised analytes <strong>and</strong> large MW analytes a , but theimpact on n<strong>eu</strong>tral molecules is typically smallerCurrent discussions / theory focus on changes inanalyte molar volume as the principle cause forchanges in k <strong>and</strong> observedSuccessful HPLC ↔ UHPLC possible...the analyst justneeds to be vigilanta Fallas et al. / J. Chromatogr. A 1209 (2008) 195–205

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!