8. AppeNdIx<br />

1C2f Format<br />

The AES3 standard includes the following two options for 96 kHz<br />

sample rate operation:<br />

• To double the frame rate from the previous 48 kHz to 96 kHz<br />

(not possible for older equipment). This is the normal operation<br />

of the <strong>Digilyzer</strong>.<br />

• Using the two sub-frames (two channels) of a 48 kHz frame<br />

rate AES3 signal to carry consecutive samples of a mono signal<br />

resulting in a 96 kHz sample rate stream. The samples of one<br />

96 kHz signal are packed “interleaved” into two 48 kHz signals.<br />

This allows older equipment, which transmitters and receivers<br />

are not rated for 96 kHz frame rate operation, to handle 96 kHz<br />

sample rate information.<br />

This mode is called “single channel double frequency mode” or<br />

“double wire mode” (two AES3 cables are needed for stereo).<br />

The <strong>Digilyzer</strong> also enables measurements in this mode. Just<br />

select the input format 1C2f and the input channel menu will<br />

indicate “A i B” (or eg. “1 i 2” for ADAT). This indicates that<br />

channel A and B are used in an interleaved manner (1C2f<br />

mode).<br />

Appendix<br />

Fig 8-01, 1C2f selection<br />

The <strong>Digilyzer</strong> is specified for samples rates up to 96 kHz, so for<br />

a 1C2f signal the resulting sample frequency shall not exceed<br />

this value.<br />

53<br />

A and B are<br />

interleaved