Section 5 <strong>Scanner</strong> ® <strong>2000</strong> microEFM<br />
MVT Replacement<br />
Important Press the ENTER/SAVE key on the keypad before disconnecting the battery to save accumulated<br />
flow run and turbine volume totals (grand total and current day total), and<br />
energy and mass totals to memory.<br />
To replace the MVT of the <strong>Scanner</strong> <strong>2000</strong>, perform the following steps:<br />
1. Unscrew the cover of the enclosure counter-clockwise until it separates from the main body of the enclosure.<br />
2. Using a small standard blade screwdriver, remove the two #4-40 × 7/8” screws located to the right and<br />
leftside of the display (Figure 5.1, page 76).<br />
3. Lift the display/keypad assembly from the enclosure.<br />
4. Unplug the battery cable from connector J1 on the circuit board.<br />
5. Disconnect the sensor ribbon cable from the J5 connector on the circuit board as follows:<br />
80<br />
a. Lift the latch from the black clip securing the ribbon cable (Figure 5.2, page 77).<br />
b. When the latch is fully open, the ribbon cable will release freely.<br />
6. Loosen the set screw in the side of the MVT adapter.<br />
7. Rotate the adapter counterclockwise to break the connection with the MVT sensor body.<br />
8. Detach the MVT sensor from the adapter, pulling the ribbon cable free.<br />
9. Remove the replacement MVT from its packaging and route the ribbon cable through the adapter and up<br />
into the <strong>Scanner</strong> <strong>2000</strong> enclosure.<br />
10. Screw the MVT into the adapter until it meets with resistance.<br />
11. Slowly unscrew the MVT sensor until the vents on the sides of the MVT are oriented to the back of the<br />
enclosure.<br />
12. Replace the set screw in the adapter and tighten.<br />
13. Connect the ribbon cable from the sensor to the MVT connector (J5) on the main circuit board.<br />
14. Reconnect the battery cable to connector J1 on the main board.<br />
15. Reinstall the display/keypad assembly in the enclosure, using the screws that were removed in step 2.<br />
16. Recalibrate the <strong>Scanner</strong> <strong>2000</strong> and replace the cover on the enclosure.<br />
Important Do not overlook the need to recalibrate the <strong>Scanner</strong> <strong>2000</strong>. MVTs that are shipped independently<br />
of a <strong>Scanner</strong> <strong>2000</strong> are not calibrated to compensate for atmospheric pressure;<br />
therefore, a <strong>Scanner</strong> <strong>2000</strong> will not display accurate pressure readings until it is recalibrated.