<strong>Scanner</strong> ® <strong>2000</strong> microEFM Section 5<br />

18. Recalibrate the <strong>Scanner</strong> <strong>2000</strong> and replace the enclosure cover.<br />

Important Do not overlook the need to recalibrate the <strong>Scanner</strong> <strong>2000</strong>. Boards that are shipped independently<br />

of a <strong>Scanner</strong> <strong>2000</strong> are not calibrated to compensate for atmospheric pressure;<br />

therefore, a <strong>Scanner</strong> <strong>2000</strong> will not display accurate pressure readings until it is recalibrated.<br />

19. Re-establish power to the peripheral circuitry.<br />

Keypad Replacement<br />

!<br />

WARNING: To prevent ignition of hazardous atmospheres, do not remove the cover while circuits<br />

are alive. The <strong>Scanner</strong> <strong>2000</strong> poses no hazard when opened in a safe area.<br />

To replace the keypad of the <strong>Scanner</strong> <strong>2000</strong>, perform the following steps:<br />

1. Unscrew the cover of the enclosure counter-clockwise until it separates from the main body of the enclosure.<br />

2. Using a small standard blade screwdriver, remove the two #4-40 × 7/8” screws located to the right and<br />

left side of the display (Figure 5.1, page 76).<br />

3. Lift the display/keypad assembly from the enclosure.<br />

4. Remove the two #4-40 × 5/16” screws fastening the circuit assembly to the keypad (Figure 5.3, page 78).<br />

5. Disconnect the keypad ribbon cable from the J7 connector on the LCD side of the circuit assembly as follows:<br />

a. Grasp the black clip between a thumb and forefinger (Figure 5.4, page 78).<br />

b. Squeeze both sides of the clip and gently pull to release the clip from the plastic connector that holds<br />

it in place. DO NOT PULL on the ribbon cable. When the black plastic clip is properly disengaged,<br />

the cable will release freely.<br />

6. Remove the old keypad.<br />

7. Connect the ribbon cable of the replacement keypad to the J7 connector on the LCD side of the circuit<br />

assembly as follows:<br />

a. Insert the end of the ribbon cable into the plastic clip.<br />

b. While holding the ribbon cable in place, press the black plastic clip into the connector until it snaps.<br />

8. Mount the circuit assembly to the keypad with the two #4-40 × 5/16” screws removed in step 4.<br />

9. Mount the display/keypad assembly to the enclosure with the two #4-40 × 7/8” screws removed in step 2.<br />

10. Recalibrate the <strong>Scanner</strong> <strong>2000</strong> if necessary.<br />

11. Replace the enclosure cover and tighten.<br />