22.11.2012 Views

Scanner 2000 Steam Mass Flow Transmitter ... - Spirax Sarco

Scanner 2000 Steam Mass Flow Transmitter ... - Spirax Sarco

Scanner 2000 Steam Mass Flow Transmitter ... - Spirax Sarco

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Scanner</strong> ® <strong>2000</strong> microEFM Section 5<br />

Important If possible, download the configuration settings and all archive logs before replacing the<br />

circuit board. Press the ENTER/SAVE key on the keypad before disconnecting the battery<br />

to save accumulated flow run and turbine volume totals (grand total and current day<br />

total), and energy and mass totals to memory.<br />

To replace the circuit assembly, perform the following steps:<br />

1. Unscrew the cover of the enclosure counter-clockwise until it separates from the main body of the enclosure.<br />

2. Using a small standard blade screwdriver, remove the two #4-40 × 7/8” screws located to the right and<br />

left side of the display (Figure 5.1, page 76).<br />

3. Lift the display/keypad assembly from the enclosure.<br />

4. Record the locations of all cable connections to the circuit board.<br />

5. Using a small standard blade screwdriver, remove all wiring from terminal blocks TB1, TB2, and TB3,<br />

and J2, ensuring that all wiring that is connected to powered circuits is insulated with tape.<br />

6. Unplug the battery cable from connector J1 on the circuit board.<br />

7. Disconnect the sensor ribbon cable from the J5 connector on the circuit board as follows:<br />

a. Lift the latch from the black clip securing the ribbon cable (Figure 5.2).<br />

b. When the latch is fully open, the ribbon cable will release freely.<br />

Figure 5.2—Latch securing the ribbon cable<br />

8. Remove the original circuit board/keypad assembly from the enclosure.<br />

9. Remove the two #4-40 × 5/16” screws fastening the circuit board to the keypad (Figure 5.3, page 78).<br />

10. Remove the keypad ribbon cable from the J7 connector on the LCD side of the circuit board by pressing<br />

in on the sides of the black plastic clip and pulling gently on the clip. Do not pull on the ribbon cable; the<br />

cable will release freely when the clip opens (Figure 5.4, page 78).<br />

11. Remove the circuit board from the enclosure and remove the replacement circuit board from its packaging.<br />

12. Connect the keypad ribbon cable to the J7 connector on the LCD side of the new circuit board by sliding<br />

the end of the ribbon into the black clip as far as it will go and pressing the black plastic clip into the connector<br />

until it snaps.<br />

77

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!