Appendix C <strong>Scanner</strong> ® <strong>2000</strong> microEFM<br />

C-6<br />

Register<br />

(Decimal)<br />

Communications Configuration<br />

Register<br />

(Hex) Description<br />

1100 44C Port 1 - Port Usage<br />

0 - Slave<br />

1 - Master<br />

1101 44D Port 1 Slave Address<br />

[1 to 65535, excluding 252 to 255 and 64764]<br />

1102 44E Port 1 - Baud Rate<br />

0 - 300<br />

1 - 600<br />

2 - 1200<br />

3 - 2400<br />

4 - 4800<br />

5 - 9600<br />

6 - 19200<br />

7 - 38400<br />

8 - 57600<br />

9 - 115200<br />

1103 44F Port 1 - Bus Delay<br />

mS of delay before transmitting data<br />

1104 450 Port 1 - Bus Timeout<br />

mS of delay before resetting communications<br />

1105 451 Port 2 - Port Usage<br />

0 - Slave Only<br />

1106 452 Port 2 Slave Address<br />

[1 to 65535, excluding 252 to 255 and 64764]<br />

1107 453 Port 2 - Baud Rate<br />

0 - 300<br />

1 - 600<br />

2 - 1200<br />

3 - 2400<br />

4 - 4800<br />

5 - 9600<br />

6 - 19200<br />

7 - 38400<br />

8 - 57600<br />

9 - 115200<br />

1108 454 Port 2 - Bus Delay<br />

mS of delay before transmitting data<br />

1109 455 Port 2 - Bus Timeout<br />

mS of delay before resetting communications<br />

Register<br />

(Decimal)<br />

Real Time<br />

Register<br />

(Hex) Description<br />

Data<br />

Type Access Default<br />

U16 R/W 0<br />

U16 R/W 1<br />

U16 R/W 5<br />

U16 R/W 10<br />

U16 R/W 50<br />

U16 R/W 0<br />

U16 R/W 1<br />

U16 R/W 5<br />

U16 R/W 10<br />

U16 R/W 50<br />

Data<br />

Type Access<br />

1200 4B0 Year (Real Year = register value plus <strong>2000</strong>) U16 R/W<br />

1201 4B1 Month [1-12] U16 R/W<br />

1202 4B2 Day [1-31] U16 R/W<br />

1203 4B3 Hour [0-23] U16 R/W<br />

1204 4B4 Minute [0-59] U16 R/W<br />

1205 4B5 Second [0-59] U16 R/W<br />

This block of registers is used to set the instrument’s internal clock. To set the time, it is recommended that all<br />

registers be written in one message.<br />

The time and date can also be read in the holding register groups as floating-point data.