For service assistance, please call your authorized <strong>ActiveCare</strong> dealer.Table 2: <strong>Scooter</strong> Controller Internal DiagnosticsFlashcodePossible CauseSolution1 The battery needs charging.2 Battery voltage is too low.3 Battery voltage too high4 Current limit time outThe battery voltage has dropped below23.3 volts in neutral. Rechargebatteries.The battery has dropped past 16.5volts and is not sufficient to allow safedriving. Check the battery conditionand the connections.The battery has exceeded 32 volts.Check the battery condition and theconnections.The motor current has reached toohigh a value. Check the condition ofthe motor and loom. Contact yourService Agent.5 Park brake fault Check park brake conditions.6 Speed control out of neutral7 Speed control faultReturn speed pod to neutral and restsystem. Readjust the speed lever toneutral if necessary. Contact yourService Agent.Check speed pod wiring for open orshort circuits. Check speed pot set-up.Contact your Service Agent.8 Motor fault Contact your Service Agent.9 Internal (controller) fault Contact your Service Agent.19 <strong>Osprey</strong> 4410 Owner’s <strong>Manual</strong>www.<strong>ActiveForever</strong>.com