Troubleshooting<strong>Proline</strong> <strong>Promass</strong> 83 <strong>Modbus</strong> <strong>RS485</strong><strong>Modbus</strong> No. Device status message(local display)CauseRemedy / spare partRegister:6859Data type:IntegerRegister:6821Data type:String (18 byte)114 NOISE LIM. CH0 703 P: NOISE LIM. CH0!: # 703115 NOISE LIM. CH1 704 P: NOISE LIM. CH1!: # 704Overdriving of the internalanalog to digital converter.Causes:– Cavitation– Extreme pressure pulses– High gas flow velocityA continuation of themeasurement is still possible!Change or improve process conditions, e.g. byreducing the flow velocity.116 FLOW LIMIT 705 P: FLOW LIMIT$: # 705124 ADJ. ZERO FAIL. 731 P: ADJ. ZERO FAIL!: # 731The mass flow is too high.The electronics' measuringrange will be exceeded.The zero point adjustment isnot possible or has beencanceled.Reduce flowMake sure that zero point adjustment is carried out at"zero flow" only (v = 0 m/s).Page 98 • MN0M030Issue/Rev. 0.2 (12/12)