10.07.2015 Views

Proline Promass 83E/F/O Modbus RS485 Operating Instructions

Proline Promass 83E/F/O Modbus RS485 Operating Instructions

Proline Promass 83E/F/O Modbus RS485 Operating Instructions

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Proline</strong> <strong>Promass</strong> 83 <strong>Modbus</strong> <strong>RS485</strong>Operation#Warning!"Caution!5.7.4 Current output configurationThe current output is configured as "active" or "passive" by means of various jumpers on the currentsubmodule.Risk of electric shock. Exposed components carry dangerous voltages.Make sure that the power supply is switched off before you remove the cover of the electronicscompartment.1. Switch off power supply.2. Remove the I/O board ä 102.3. Position the jumpers (see Figure).Risk of destroying the measuring device. Set the jumpers exactly as shown in the diagram.Incorrectly set jumpers can cause overcurrents that would destroy either the measuring device orexternal devices connected to it.4. Installation of the I/O board is the reverse of the removal procedure.Fig. 39:Configuring the current input with the aid of jumpers (I/O board)1 Current output1.1 Active current output (default)1.2 Passive current outputIssue/Rev. 0.2 (12/12) MN0M030 • Page 57

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!