<strong>Proline</strong> <strong>Promass</strong> 83 <strong>Modbus</strong> <strong>RS485</strong>Technical dataMeasuring ranges for gases, general, (except <strong>Promass</strong> H (Zr))The full scale values depend on the density of the gas. Use the formula below to calculate the fullscale values:g max(G) = g max(F) (G) : x [kg/m (lb/ft )]g max(G) = Max. full scale value for gas [kg/h (lb/min)]g max(F) = Max. full scale value for liquid [kg/h (lb/min)] (G) = Gas density in [kg/m (lb/ft )] for process conditionsHere, g max(G) can never be greater than g max(F)Measuring ranges for gases (<strong>Promass</strong> F, O):DNx[mm][inch]8 3/8 6015 ½ 8025 1 9040 1½ 9050 2 9080 3 110100 4 130150 6 200250 10 200Measuring ranges for gases (<strong>Promass</strong> E)DNx[mm][inch]8 3/8 8515 ½ 11025 1 12540 1½ 12550 2 12580 3 155Measuring ranges for gases (<strong>Promass</strong> P, S, H (Ta))DNx[mm][inch]8 3/8 6015 ½ 8025 1 9040 1) 1½ 1) 9050 1) 2 1) 901)only <strong>Promass</strong> P, SIssue/Rev. 0.2 (12/12) MN0M030 • Page 109