6 Patient PreparationReusable ECG Sensors!CautionBaseline drift. Never mix patient electrode types or brands. Dissimilarmetals or other incompatibilities may cause considerable baseline driftand may increase trace recovery time after defibrillation.ElectrolyteNOTE: Watches or jewelry may be damaged by contactwith electrolyte.Cardiac Science Corporation recommends using Signa gel ® electrode gelwith reusable sensors because it provides excellent conductivity betweenthe skin and sensor.Reusable sensors (Welsh bulbs and limb plates) must be kept clean. Dirt orelectrolyte build-up may degrade ECG quality:• Wash after each use (use a mild dish soap if necessary)• Never use a metallic pad to clean plates and Welsh bulbs (toavoid damaging the plating)• Discard if corroded.Applying limb plates!CautionPatient skin irritation. Limb Plates may contain nickel, a material that cancause skin irritation. Monitor the electrode site and, if irritation occurs,use an alternate electrode.1. Expose the arms and legs of the patient.2. Connect sensor straps to the “ears” of the limb plates.6-6 <strong>Atria</strong> Electrocardiograph Operating Instructions