6 Patient PreparationPatient PreparationThe following steps are necessary for patient comfort and to minimizemuscle artifacts:• Assure the patient there is no danger or discomfort involved.• Explain that full cooperation will produce a valuablediagnostic record.• Ensure the patient is comfortable and relaxed.- Lay the patient on a cot or padded table large enoughto support both arms and legs.- Arms should rest at the side and the legs should lieflat and not touching.- Support the head with a pillow.- Avoid cold drafts and cover the patient with a blanketto prevent shivering (leave the chest and sensor sitesexposed).Skin PreparationRefer to the “Resting ECG Lead Placement & Coding Chart” on pg. 6-4 fordetails on where sensor sites are located.For information on alternate chest lead and other lead placements, see“Alternative Lead Placements” on pg. 6-8.1. Use the thumb and forefinger to part hair when necessary. Clipping orshaving may be required.2. If skin has lotion on it or the patient is diaphoretic, clean skin withalcohol and abrade slightly with a dry gauze.3. Ensure the skin is completely dry.6-2 <strong>Atria</strong> Electrocardiograph Operating Instructions