5Care and maintenanceInstruments319.240 Cleaning Brush 2.9 mm,for Cannulated Instruments519.970 Oil Dispenser with Synthes Special Oil,50 ml321.158 Combination Wrench 8.0 mm398.759 Combination Wrench 13 mmBefore cleaning and sterilization, remove any attachmentfrom the sliding mechanism (arm, reduction attachment andspiked disk) and remove the synthetic screens from thesliding mechanism. Clean the cannulated feed rod manually,using a 2.9 mm cleaning brush.Important: Lubricate the movable parts (e.g. clamp springof the sliding mechanism, couplings of attachment arms) ofthe <strong>Collinear</strong> <strong>Reduction</strong> <strong>Clamp</strong> after each use with autoclavableSynthes oil. After oiling, push the button on the attachmentcoupling and slide the feed rod several times toensure proper distribution of the oil.Before cleaning and sterilization, close the reductionattachment either by hand or by using the combinationwrench 8 mm (321.158) and the combination wrench 13 mm (398.759).Synthes 9