2818.2 TEMPERATURE SETTING FOR THE COOLERThe pre-cooling stage is set at the factory to a regulated temperature of +1°C.RUrefCurrent-Temp.P3+ °C -UtempUrefR i > 100kC9Setting-Temp.Connect a d.c. voltmeterto terminals 1, 2 and 3!Figure 11 Temperature adjustmentSetting of the regulated temperature is done by trimming potentiometer P3, on the EC automaticcontrol board of the cooler. The setting range covers temperatures from 0°C to 20°C. Turning it to theright sets a lower temperature, and turning it to the left sets a higher temperature.Connecting an external voltmeter to plugs 2 and 3 allows the nominal set temperature to be read offand controlled. A voltage value of 0.1V corresponds to a temperature of 1°C.It is possible to measure and control the current temperature at measuring sockets 1 and 2.NOTE!Freezing up in the pre-cooling stage will endanger the operation of thecooler unit. For this reason the cooler temperature must never fall below+1°C!19 TEMPERATURE SENSORSThe pre-cooling stage temperature sensor is a PT100 element. STP 35 temperature sensors areavailable for the low temperature stages and .19.1 CHECKING THE TEMPERATURE SENSOR OF THE PRE-COOLING STAGEThere are two methods for checking the PT100 element, as follows:Voltage methodIn order to check the sensor for the cooler currently in operation, the actual voltage at the measuringsockets 1 and 2 of the EC automatic control board must be measured as per section 18.2 above.<strong>Gas</strong> sampling and gas conditioning technology3-4.20-ME

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!