The heat energy from the cooling system is drawn off by the forced ventilated condenser. The LEDfunction display on the front of the cooler shows the operating status. Alarm warnings for over- andunder-temperature are given as a collective status alarm via a relais output with two potential-freechangeover contacts.1411 RECEPTION AND STORAGEThe EC-30(/FD) gas cooler is a complete pre-installed unit.Carefully inspect the EC-30(/FD) and any special accessories included immediately on arrival byremoving them from the packing and checking for missing articles against the packing list!Check the items for any damage in transit and, if required, inform the shipping insurancecompany immediately of the damage found!The tubing for the heat exchangers is dismounted for shipping and it is separately inside of thepackage.NOTE!NOTE!The cooler must be stored in a weather-protected frost-free area!During transport and when in storage, the cooler has always to stand verticalwith the transport feet positioned underneath so that the oil in the closedcompressor circuit cannot run out of this compressor case. If the cooler istransported on its back by mistake, it has to stand in the operating positionfor approx. 24 hours before it is switched on!<strong>Gas</strong> sampling and gas conditioning technology3-4.20-ME