Message Priority Cause Action/Concerns RepairSystemLeak?VentilateManuallyVolumeApneaVol Apnea> 2 minV T Comp OffLowHighMediumHighMedium(low afteracknowledge)Leak detectedbetween ventilatorand patient circuit.A severemalfunctionpreventsmechanicalventilation andmonitoring.Other alarms mayalso occur.No mechanicalbreaths orspontaneousbreaths > 5 mL inlast 30 seconds.No mechanicalbreaths orspontaneousbreaths > 5 mL inlast 120 seconds.The system suppliesthe set breath butcannot adjustventilation forcompliance andresistance losses,etc.Look for leaks inthe absorbersystem.Problem with flowsensors?Ventilate manually.Use a stand-alonemonitor.Cycle system power(On- Standby-On).If the alarm clears,restart mechanicalventilation.Check patient.Bag as needed.Check fordisconnects.If the patient is on aheart lungmachine, selectCardiac Bypass onthe alarm menu.Check patient.Bag as needed.Check fordisconnects.If the patient is on aheart lungmachine, selectCardiac Bypass onthe alarm menu.Adjust V T manuallyand continuewithoutcompensation, orchange to thepressure mode.In pressure modeset P insp.Calibrate theflow sensors.Drain waterbuildup from thebreathingsystem andinspect for leaks(repair).Inspect orreplace flowsensors.Contact aqualified servicerepresentative.- - -- - -Inspect the twoflow sensors.22 S/5 <strong>Aespire</strong> Quick Reference Guide