Download Pccr 1417 Acl4 System - Genesis Digital Imaging
Download Pccr 1417 Acl4 System - Genesis Digital Imaging
Download Pccr 1417 Acl4 System - Genesis Digital Imaging
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
30PcCR <strong>1417</strong>ACL4 <strong>System</strong> Users GuideFigure 12: The Setup Dialog BoxAdjusting the <strong>System</strong> to an X-ray UnitThe Scanner Interface is designed to work with up to four different x-ray units. A unitcan be assigned to different X-ray machines (e.g., mobile and stationary) or to widelydifferent x-ray settings (e.g., premature babies and pregnant women in a maternity wing)on the same unit. Each unit is represented by a number, and includes a unique set ofPM Gain (and offset) values, and filtration values for each sub-organ. Before x-raying,select the machine you are going to use by clicking that unit’s number.Each unit can be given a name on the Anatomical tab. After you name a unit, thatname appears on the main dialog whenever that unit is activatedTo adjust the system to an x-ray unit:1. In the Scanner Interface (Figure 13) click the desired unit (1, 2, 3, or 4).This can also be done on the Anatomical tab of the Setup dialog.