Gas Chromatography (GC) (IUPAC Compendium of Chemical Terminology):

27.06.2015 Views

Lecture 3. Gas chromathography. Split Injection. Used for samples with analyte concentration > 0.1 %. Only 0.2 – 2 % of the sample is delivered to column. Injector temperature is high, e.g. 350 ºC. 102 ml/min 1 ml/min 100 ml/min 1ml/min The sample is injected rapidly through the septum into evaporation zone. The injector temperature is kept high to promote fast evaporation. A brisk flow of the carrier gas sweeps the sample through the mixing chamber. At the split point, small fraction of vapors enters the column but most passes to waste vent. Split ratio (the proportion of the sample that does not reach the column) is typically 50:1 to 600:1. Septum purge gas flow: prevents the column during injection and chromatography from hot rubber septum gases and the excess of the sample vapors. ● ● ● Advantages of split injection: narrow solute peaks; suitable for qualitative analysis; minimize the solvent effect. Drawbacks: ● ● ● requires rather high concentration of analyte; split ratio makes the quantitative analysis more complex; not suitable for very expensive or toxic compounds. 22

Lecture 3. Gas chromathography. Splitless injection. Used for traces analysis with analyte concentration < 0.01%. Volume of solution injected is ~ 2 μl. Injection time is ~ 2 sec (SLOW INJECTION). Injector temperature ~220 ºC. ~ 80% of the sample is applied to the column. 2 ml/min 0 ml/min 1 ml/min 1ml/min Solvent trapping. The initial column temperature is set 40 ºC below the boiling point of the solvent. Therefore the solvent condenses in the beginning of the column and traps the analyte to produce a narrow plug in the beginning of the column. For solvent trapping, the analyte concentration should be < 0.01%. Cold trapping. The initial column temperature is 150 ºC than the boiling points of analytes of interest. Solvent and low-boiling components are eluted rapidly, whereas high-boiling solutes remain as narrow band. The column is then rapidly warmed to initiate chromatography of highboiling solutes. Stationary-phase film thickness must be ≥2 μm. Cryogenic focusing is the variation of cold trapping for low-boiling solutes. The column is initially cooled with N 2 or CO 2 . ● ● ● ● ● Advantages of splitless injection: suitable for quantitative and qualitative analysis; narrow peaks of analyte. Drawbacks: broad solvent peak; retention times depend on solvent evaporation speed; solvent affects the shape of the peaks. 23

Lecture 3. <strong>Gas</strong> chromathography.<br />

Splitless injection.<br />

Used for traces analysis with analyte concentration < 0.01%.<br />

Volume <strong>of</strong> solution injected is ~ 2 μl.<br />

Injection time is ~ 2 sec (SLOW INJECTION).<br />

Injector temperature ~220 ºC.<br />

~ 80% <strong>of</strong> the sample is applied to the column.<br />

2 ml/min<br />

0 ml/min<br />

1 ml/min<br />

1ml/min<br />

Solvent trapping. The initial column temperature is set 40 ºC below the boiling point <strong>of</strong> the<br />

solvent. Therefore the solvent condenses in the beginning <strong>of</strong> the column and traps the analyte<br />

to produce a narrow plug in the beginning <strong>of</strong> the column. For solvent trapping, the analyte<br />

concentration should be < 0.01%.<br />

Cold trapping. The initial column temperature is 150 ºC than the boiling points <strong>of</strong> analytes <strong>of</strong><br />

interest. Solvent and low-boiling components are eluted rapidly, whereas high-boiling solutes<br />

remain as narrow band. The column is then rapidly warmed to initiate chromatography <strong>of</strong> highboiling<br />

solutes. Stationary-phase film thickness must be ≥2 μm.<br />

Cryogenic focusing is the variation <strong>of</strong> cold trapping for low-boiling solutes. The column is<br />

initially cooled with N 2 or CO 2 .<br />

●<br />

●<br />

●<br />

●<br />

●<br />

Advantages <strong>of</strong> splitless injection:<br />

suitable for quantitative and qualitative analysis;<br />

narrow peaks <strong>of</strong> analyte.<br />

Drawbacks:<br />

broad solvent peak;<br />

retention times depend on solvent evaporation speed;<br />

solvent affects the shape <strong>of</strong> the peaks.<br />

23

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!