General Description<br />

Windows-Based Software<br />

a. Calibrate<br />

This function is used to calibrate the monitor. Invoke the calibration<br />

routine by typing > C or > ENTER if “Calibrate” is highlighted.<br />

The operator is prompted for the portal serial number and the<br />

operator ID. The operator ID can be an alphanumeric string of up<br />

to 16 characters. After the responses have been entered, the source<br />

selection, monitor serial number, and operator ID are echoed to the<br />

screen for verification. If incorrect, the operator can edit the<br />

information at this time by first pressing > F1. Striking any other<br />

key accepts the information and presents the operator with another<br />

submenu as shown in Figure 5.2.<br />

04:01:21<br />

Establish PC/<strong>PM</strong>-7 Communications<br />

PASSWORD OK<br />

>Connect >Disk Edit >Reports >Password >Quit<br />

<strong>PM</strong>-7 #1<br />

Calibrate the Connected <strong>PM</strong>-7<br />

>Calibrate >Utilities >Edit >Quit<br />

Run High Voltage Plateaus<br />

>Plateaus >Efficiencies >Shield Factors >Quit<br />

Figure 5.2. Thermo Fisher Scientific <strong>PM</strong>-7 Calibration Program<br />

i. If “Plateaus” is selected, the first prompt that appears allows the<br />

operator to identify the calibration source that will be used. The<br />

source can be selected from a list of up to four sources that can<br />

be defined in the System Parameter file. A composite source<br />

made up of two sources from that list can be identified or a<br />

source not identified in the list can be entered. The <strong>PM</strong>-7<br />

corrects the activity for decay each time calibration is performed.<br />

The computer will first run a background plateau, then a source<br />

plateau, and finally, will calculate figures of merit for each<br />

detector at each highvoltage setting. The range of voltages and<br />

the size of each increment in high voltage is set in the system<br />

parameters. The area should be clear of all radioactive sources<br />

Thermo Fisher Scientific<br />

Personnel Monitor -7 Techical Guide_5-5_