General Description<br />
Windows-Based Software<br />
a. Calibrate<br />
This function is used to calibrate the monitor. Invoke the calibration<br />
routine by typing > C or > ENTER if “Calibrate” is highlighted.<br />
The operator is prompted for the portal serial number and the<br />
operator ID. The operator ID can be an alphanumeric string of up<br />
to 16 characters. After the responses have been entered, the source<br />
selection, monitor serial number, and operator ID are echoed to the<br />
screen for verification. If incorrect, the operator can edit the<br />
information at this time by first pressing > F1. Striking any other<br />
key accepts the information and presents the operator with another<br />
submenu as shown in Figure 5.2.<br />
04:01:21<br />
Establish PC/<strong>PM</strong>-7 Communications<br />
PASSWORD OK<br />
>Connect >Disk Edit >Reports >Password >Quit<br />
<strong>PM</strong>-7 #1<br />
Calibrate the Connected <strong>PM</strong>-7<br />
>Calibrate >Utilities >Edit >Quit<br />
Run High Voltage Plateaus<br />
>Plateaus >Efficiencies >Shield Factors >Quit<br />
Figure 5.2. Thermo Fisher Scientific <strong>PM</strong>-7 Calibration Program<br />
i. If “Plateaus” is selected, the first prompt that appears allows the<br />
operator to identify the calibration source that will be used. The<br />
source can be selected from a list of up to four sources that can<br />
be defined in the System Parameter file. A composite source<br />
made up of two sources from that list can be identified or a<br />
source not identified in the list can be entered. The <strong>PM</strong>-7<br />
corrects the activity for decay each time calibration is performed.<br />
The computer will first run a background plateau, then a source<br />
plateau, and finally, will calculate figures of merit for each<br />
detector at each highvoltage setting. The range of voltages and<br />
the size of each increment in high voltage is set in the system<br />
parameters. The area should be clear of all radioactive sources<br />
Thermo Fisher Scientific<br />
Personnel Monitor -7 Techical Guide_5-5_