Chapter 4 Algorithms<br />
Counting Algorithms<br />
The sum channel determines the minimum time required to measure the<br />
specified RDA at the required confidence level while maintaining the<br />
selected false alarm probability. That time is calculated as follows:<br />
T = SF R B + z ( RDA × eff) + R<br />
------------------------------------------------------------------------- B<br />
RDA × eff<br />
Where:<br />
“T” Is the count time, in seconds.<br />
2<br />
Eq. 2.1<br />
“SF” Is the SIGMA FACTOR which sets the false alarm rate. “SF” is the<br />
multiplier of the standard deviation of the average background count<br />
rate.<br />
“R B ” Is the sum channel's average background count rate in counts per<br />
second.<br />
“z” Determines the confidence level, or probability of detection. z is the<br />
multiplier of the standard deviation of the combined average<br />
background and RDA count rate.<br />
“RDA” Is the “Instrument RDA” or sum channel's Reliably Detectable<br />
Activity expressed in units of dps.<br />
“eff” Is the 4-pi efficiency of the sum channel computed as the ratio of<br />
net count rate in counts per second over the decay-corrected source<br />
activity in dps. The efficiency of the sum zone should be measured with<br />
a calibration source of the isotope of interest placed in the center of the<br />
portal.<br />
If the calculated count time in the sum channel exceeds the selected<br />
maximum count time, a high background alarm will occur.<br />
Thermo Fisher Scientific Personnel Monitor -7 Techical Guide_ 4-1