General Description<br />

Detectors<br />

Detectors<br />

Six 1.5 inch (3.81 cm) thick plastic scintillation detectors surround the<br />

portal opening. The head and foot detectors are 11 inches x 19 inches (27.9<br />

cm x 48.3 cm), and the side detectors are 11 inches x 38 inches (27.9 cm x<br />

96.5 cm).<br />

Typical radiological responses of the side detectors are:<br />

• 137 Cs at 3 inches: 2.9 cps/nCi (7.4%).<br />

• 60 Co at 3 inches: 5.8 cps/nCi (14.8%).<br />

• Omnidirectional background: 100 cps/mR/h (detector behind<br />

cover, without shielding).<br />

Audible Annunciators<br />

Several audible indicators are utilized to simplify portal operation:<br />

• A short tone indicates that a user has started a count cycle by<br />

entering the portal. A user can disable this feature. To disable open<br />

SW2-4 on master annunciator board. Reference drawing<br />

1468-C109 is section 5 and 17000-A61.<br />

• A continuous tone indicates that contamination has been detected.<br />

• An intermittent tone indicates that a user exited the portal before<br />

the count cycle was complete.<br />

• A chime indicates the completion of a count cycle with no<br />

contamination detected. A user can disable this feature if the<br />

sensitivity is such that the instrument will support walk-through<br />

mode. To disable, open SW2-1 on the Master Controller Board.<br />

Refer to Switch Setting drawing No. 11468-C109 in Section 5, also<br />

see 17000-A61.<br />

Operational<br />

Parameters<br />

The <strong>PM</strong>-7 utilizes several operational parameters which can be set directly<br />

by the user or indirectly using the calibration program:<br />

• OVERALL RDA (nCi) is the portal sum zone's reliably detectable<br />

activity (RDA).<br />

• OVERALL EFFICIENCY (%) is the portal sum zone's 4-pi<br />

efficiency to a source placed at its center.<br />

• SIGMA FACTOR controls the portal's false alarm rate. The sum<br />

zone and each of the detector sigma factors may be set individually.<br />

Thermo Fisher Scientific<br />

Personnel Monitor -7 Techical Guide_3-3_