Section 1<br />

()ducti n<br />

General Description<br />

The Eberline model <strong>PCM</strong>-2 is a monitoring instrument which provides rapid and<br />

convenient detection and localization of alpha and beta/gamma contamination on<br />

persoill1el. Sixteen (16) gas proportional detectors, designed for rapid replacement<br />

and/or repair, provide superior localization of contamination and background<br />

rejection. A contoured array of 34 counting zones, each with separate alpha and<br />

beta/gamma channels, provides data from which the instrument can determine the<br />

presence of both localized and distributed contamination. In addition, up to 75<br />

"Sum Zones" can be defined, each comprised of 2,3 or 4 adjacent detectors,for a<br />

maximum detection of contamination that is spread over two or more detectors.<br />

A "Sum Channel" is the summation of all 34 detectors into a single counting<br />

channel. With ex and i3/'Y discrimination, the 34 discrete channels, 75 Sum Zones,<br />

and the Sum Charmel provide a total of 220 measurements per count cycle. Since<br />

a person is measured with two count cycles, there are 440 independent<br />

measurements performed per person. The"Sum Channel" is thereby capable of<br />

detecting low-level, widely-distributed contamination.<br />

Individual detector channels within the <strong>PCM</strong>-2 are independently controlled by<br />

distributed microprocessors. In addition, a 20 Mhz 80386SX based computer is<br />

built in to provide a user-friendly interface for the system. This enhanced<br />

controller also simplifies calibration and maintenance of the unit and presents test<br />

results to the user in a clear and easily understood graphic format. A full-size<br />

keyboard is stored internal to the unit for use during setup, calibration and<br />

troubleshooting. The <strong>PCM</strong>-2 is also capable of logging measurement data in any<br />

of several formats to a printer and/or host computer system.<br />

The <strong>PCM</strong>-2 features an functionally ergonomic design which maximizes user body<br />

contact and minimizes dead areas between detectors. A color VGA monitor is<br />

used to display information during normal operation as well as during setup,<br />

calibration and maintenance. If contamination is defected during a measurement<br />

cycle a color graphic image is presented to indicate contamination location(s).<br />

<strong>PCM</strong>2.MANI REV AI April 1995<br />