08.06.2015 Views

PCM-2 Manual.pdf - Voss Associates

PCM-2 Manual.pdf - Voss Associates

PCM-2 Manual.pdf - Voss Associates

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

• Detector channels whose corrected average background count rate falls within<br />

the tolerance band are considered to be normal. If a detector's alpha or beta<br />

average background count rate exceeds the upper limit of the tolerance band,<br />

that detector is flagged with a HI SENSITIVITY FAILURE. A LOW<br />

SENSITIVITY FAILURE occurs whenever a detector channel falls below the<br />

lower limit.<br />

The environment that a <strong>PCM</strong>-2 operates in can cause sensitivity failures even<br />

though all detectors can be operating properly. For that reason, the first<br />

recommended step is to analyze the Geometry Factors and Sensitivity Factors and<br />

make changes to those parameters as necessary.<br />

• Default Geometry Factors are set in inverse proportion to the detectors area,<br />

using detector #O's area as the benchmark. Factors other than a detector's<br />

size can contribute to its cout rate, e.g., a streaming background or a shield<br />

that affects only part of the instrument. If changing Geometry Factors is<br />

warranted, observe the background averages in the VIEW menu -- insure that<br />

the Correction is toggled off -- and assess relative values of count rates to<br />

establish appropriate Geometry Factor values. Geometry Factors are edited in<br />

the Detector Parameter screen under the EDIT menu.<br />

• Default sensitivity factors are 0.50 for the beta channels and 0.10 for the<br />

alpha channels. While these values will work for many installations, an<br />

environment that involves wide fluctuations of background accompanied by<br />

directional characteristics may warrant lowering the sensitivity factors.<br />

Sensitivity factors are established in the System Parameters screen under the<br />

EDIT menu.<br />

If count rates are high or low enough to indicate a true sensitivity failure, the<br />

following is suggested.<br />

• A low sensitivity failure may be caused by counting gas starvation. Verify<br />

positive gas supply, correct pressure and flow rate settings and search out any<br />

leaks or kinked hoses that may be present. Once corrected, allow sufficient<br />

purge time before placing the channel back into service.<br />

• Other possible problems will be narrowed down to a failed detector assembly<br />

or its complimentary MDB (Modular Detector Board). By exhanging the<br />

MDB with that of a known good channel it is possible to isolate the problem<br />

to either the board or the detector. The faulty assembly should be replaced.<br />

NOTE: Whenever and MDB is replaced, the DETECTOR DOWLOAD routine<br />

must be invoked to restore the correct parameters to the affected detector.<br />

High Voltage Fail<br />

Verify that the high voltage failure message was not induced by some transient<br />

condition by first executing the DETECTOR DOWNLOAD routine from the<br />

9-6<br />

<strong>PCM</strong>2.MAN/ nP.v A/April 1995

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!