exercising care to avoid detatching the taped-on nutplate. Adjust the switch<br />

body inward against the actuator arm until the switch contacts just make and<br />

then tighten the mounting screws. This should yield a sensitive switch<br />

adjustment which requires only a slight deflection of the hip switch actuator<br />

rod to trip the switch. Readjust the switch position until this sensitive<br />

switch actuation setting is acheived.<br />

On later <strong>PCM</strong>-2's (Serial no's 219 & subsequent), the hip position switch<br />

and mounting bracket are different in style and design. This improved hip<br />

switch design offers greater reliability and ease of replacement. Removal of<br />

this switch is simply a matter of removing the island's front cover panel and<br />

then removing the two flathead screws which secure the switch mounting<br />

bracket to the inner left-hand flange of the internal backing plate near the<br />

top of the island panel. This will allow the switch and its wiring, attached<br />

to the mounting bracket, to be extracted foreward through the front of the<br />

island panel. Note the relative axial mounting position of the switch plunger<br />

end to the near surface of the mounting bracket. This dimension is critical<br />

for remounting the new switch in the position required to yield the correct<br />

switch acutation sensitivity. The switch can then be removed from the<br />

mounting bracket by loosening and removing its two jam nuts and<br />

disconnecting the switch wires at the molex connector. Replacement of the<br />

new switch assembly is in the reverse order of the removal operations listed.<br />

Note that this "barrel" type switch has no audible indication of actuation and<br />

should thereby be verified for same via an electrical continuity test. Again,<br />

the proper adjustment for this hip poistion switch will result in switch<br />

contact release with only a slight displacement of the hip switch actuator<br />

rod.<br />

A kit which retrofits the old style <strong>PCM</strong>-2 hip position switch with the new<br />

improved style switch is available as Eberline part no. YPI1534592, Hip<br />

Switch Retrofit Kit.<br />

There are no other standard mechanical or electrical components in the <strong>PCM</strong>-2<br />

which require scheduled repair or replacement. Please contact Eberline Customer<br />

Service reguarding PMC-2 instrument repair or warranty repair concerns which<br />

exceed the scope of "preventative" or "routine" maintenance described herein.<br />

Reference the Warranty Repair information sheet at the front of this manual.<br />

Decontamination<br />

Cleaning<br />

The <strong>PCM</strong>-2 has been designed to allow relative ease of decontamination by<br />

providing user interface surfaces and detector panels made of stainless steel. The<br />

stainless steel material enables the use of water, detergents, and even mild<br />

solvents and lor abrasives in the decontamination process without damaging or<br />

reducing the working life of the detector panels' surfaces.<br />

Note: It is not recommended that a pressurized flow of water and/or other<br />

cleaning agents be used when cleaning the external surfaces of the instrument for<br />

9-4<br />