screws, nutplate and switch actuator assy. The switch and wiring can then<br />

be extracted from the mounting bracket where it must be unsoldered from<br />

its electrical leads. Noting the positions of the switch leads on the switch<br />

contacts, resolder the switch leads to the contacts of the new switch. Install<br />

the new switch to the switch mounting bracket in reverse order of removal.<br />

Adjust the switch within its mounting slots and/or by slightly bending the<br />

switch actuator arm so that the switch makes contact upon a slight deflection<br />

of the footplate grille (as by the weight of a user) but does not make contact<br />

with the footplate grille in place by itself. When the proper switch<br />

adjustment has been made, remount the access panel cover plate or return<br />

the unit to the upright position.<br />

Hand Position Switch<br />

The hand position switch is located lower electtronics enclosure mounted on<br />

the inside, right-hand side of the consloe base panel. It is readily accessible<br />

thm the <strong>PCM</strong>-2's side door and by removing the lower electronics enclosure<br />

cover panel. Like the weight switch, it mounts with two screws and a<br />

nutplate thm twin adjustment slots on the panel and its contacts are soldered<br />

to its electrical leads. Remove the switch and install a new switch noting to<br />

resolder the leads and to remount the switch and actuator as before. Adjust<br />

the loosely mounted switch so that the switch contacts release upon only a<br />

very slight deflection of the "floating" hand detector panel and so that the<br />

switch contacts make positively upon return of the panel back to its springloaded<br />

rest position. Tighten the switch mounting screws.<br />

Hip Position Switch<br />

This switch, which senses user hip position via displacement of the long red<br />

hip switch actuator rod, is located in the uppermost cavity of the island<br />

panel assembly above the island panel's detectors. Replacement of the<br />

switch in older <strong>PCM</strong>-2 models (Serial no's 001 thm 218) first envolves<br />

removing the island panel's front cover panel and the palm and thigh<br />

detectors from the island assembly. This affords limited accesss to the hip<br />

switch for its removal. Loosen and remove the two switch mounting screws<br />

to disassemble the switch from its mounting bracket and extract the switch<br />

and its wiring thm the rear of the island assembly. Clip or unsolder the<br />

wires from the switch and re-solder them to the new switch in the same<br />

contact locations, noting to bend the switch contacts towards the switch case<br />

as per the old switch. Tape the switch mounting nutplate to the switch body<br />

on top with a thin double-sided tape, aligning the press nuts in the nutplate<br />

with the switch's mounting holes, and puncture the tape film thm each<br />

pressnut. This will keep the mounting nutplate in place on the switch<br />

during the "blind" switch placement, mounting and adjustment operations.<br />

Route the newly prepared switch and wiring back up and into the upper<br />

island panel cavity again thm the rear of the island panel. Position the<br />

switch on the mounting bracket and loosely install the two mounting screws,<br />

<strong>PCM</strong>2.MANI REV AI April 1995 9-3