CALIBRAnON<br />

Background Plateau<br />

Background plateaus may now be run for the hand probe board if it is installed and<br />

communicating properly. Pressing F5 in the Select Detectors screen toggles the selection of the<br />

hand probe detector. If the hand probe option is installed but the board is not communicating,<br />

a message will appear in the Select Detectors screen informing the user that the hand probe is<br />

not communicating. Pressing F5 when choosing to view plateau data also selects the hand probe<br />

detector. Plateau data may be viewed even if the hand probe board has a communication failure.<br />

Source Plateau<br />

Source plateaus may now be run for the hand probe board if it is installed and communicating<br />

properly. Pressing F5 in the Select Detectors screen toggles the selection of the hand probe<br />

detector. If the hand probe option is installed but the board is not communicating, a message will<br />

appear in the Select Detectors screen informing the user that the hand probe is not<br />

communicating. Pressing F5 when choosing to view plateau data also selects the hand probe<br />

detector. Plateau data may be viewed even if the hand probe board has a communication failure.<br />

Efficiencies<br />

The hand probe detector may now be selected for an efficiency test if the hand probe is installed<br />

and is communicating properly. As for the plateaus, pressing F5 selects the hand probe detector<br />

and starts the test. If the hand probe option is installed but the board is not communicating, a<br />

message will appear in the Select Detectors screen informing the user that the hand probe is not<br />

communicating. To view the test results, press F5 to select the hand probe detector. Efficiency<br />

data may be viewed even if the hand probe board has a communication failure.<br />

EDIT<br />

Hand Probe Parameters<br />

Six new parameters are available for the hand probe. An alarm setpoint may be entered for both<br />

the alpha and the beta channels. This setpoint is expressed in the count rate units selected for<br />

the <strong>PCM</strong>-2 (CPS or CPM). In addition, a weighting factor and a background sigma factor may<br />

be entered for both the alpha and beta channels. These parameters are used in a "smoothing"<br />

algorithm which has been incorporated for the hand probe readings to allow the user to control<br />

the response time and the step change for the hand probe readings.<br />

Ref. No. 2183IMI21831N0vember 1994