False Alarm Test<br />

Calibration Report<br />

This test performs N measurements where N is selected by the operator. Each<br />

measurement is performed just as if a person was in the monitor. At the<br />

conclusion of each cycle the instrument checks for alarms and increments a<br />

counter if one or more channels have alarmed. At the conclusion of this test, the<br />

number of alarms divided by N will determine the unit's false alarm rate.<br />

Selecting this item causes the <strong>PCM</strong>-2 to generate a calibration report. If the user<br />

has not entered a serial number for the instrument other than 0000 (default value),<br />

an editing screen for the serial number will appear. The user may edit the serial<br />

number, if desired. The report will then be generated, using the new serial<br />

number, if one was entered. If a printer is available, the report may be printed<br />

for a permanent record.<br />

7-8<br />

<strong>PCM</strong>2.MAN/ Rev A/April 1995