transaction buffer, escape back to the "Status" menu and select "Overall Monitor<br />

Status". Pressing the F2 key while in this display will then clear the buffer.<br />

Print Transactions?<br />

Printer Init<br />

Day File Directory<br />

Printer Type<br />

Uses Before<br />

Background Update<br />

Background Update<br />

After Alarm?<br />

Day File<br />

Directory Storage<br />

If a printer is attached to the <strong>PCM</strong>-2 and if this field is "YES", a report of each<br />

measurement cycle is automatically printed. If this field is "ALARMS", the<br />

report is printed only after an alarm has occurred during the measurement cycle.<br />

An initialization string sent to an attached parallel printer just prior to printout of<br />

<strong>PCM</strong>-2 reports and data. NOTE: A character pitch of no less than 12 cpi is<br />

necessary for printing the <strong>PCM</strong>-2 Calibration Report.<br />

A string identifying a directory path for day file logging. See "Day File Logging"<br />

at the end of this section for more information.<br />

If a printer is connected to the <strong>PCM</strong>-2, this setting determines the set of character<br />

and control codes used. The printer type is only important when printing graphics<br />

such as high voltage plateaus.<br />

The instrument is removed from service to update background information after<br />

the specified number of measurement cycles. If there is time between users to<br />

obtain background counts, this forced update will not be required. Entering a<br />

value of zero disables this function.<br />

Setting this field to "YES" forces the <strong>PCM</strong>-2 to update its background counts after<br />

every measurement which results in an alarm. This detects contamination left on<br />

the detectors.<br />

Transactions? - Enables/disables storage of transactions in a day file if desired.<br />

Alarms only may be stored by selecting "ALARlViS".<br />

Status Changes? - Enables/disables storage of status changes in a day file.<br />

Source Check Results? - Enables/disables storage of source check results in a day<br />

file.<br />

See "Day File Logging" at the end of this section for more information.<br />

Display<br />

Midway Results?<br />

A "YES" value causes the <strong>PCM</strong>-2 to display contamination detected during the<br />

first measurement cycle. "NO" requires users to complete both measurements<br />

before any results are displayed.<br />