Count Rate Units<br />

Activity Units<br />

Count Mode<br />

10 & RWP<br />

Entry Methods<br />

Alarm Hold Time<br />

System Address,<br />

Baud Rate & Protocol<br />

Anti-Coincidence<br />

Radon Compensation<br />

Status Logging<br />

Count rates are displayed in either counts per second or counts per minute as<br />

specified by this field.<br />

The activity units are selected to be DPS, DPM, Bq or nCi.<br />

The <strong>PCM</strong>-2 may be operated in Preset All, Minimum Count Time or Fixed Count<br />

Time modes. This selection changes the format of several other display screens.<br />

User Identity and Radiation Work Permit numbers, if used, are entered via the<br />

keypad and/or an optional badge reader.<br />

The number of seconds for which an alarm sounds if contamination is detected.<br />

If the <strong>PCM</strong>-2 is connected to a host computer, these parameters determine the<br />

address of this particular instrument and the communications baud rate and<br />

protocol to be used. uLAN protocol may only be used with hardwire (RS-232,<br />

RS-485) connections. Select non-uLAN for modem connections.<br />

When anti-coincidence is enabled (default mode) particles are classified as beta or<br />

alpha on the basis of signal amplitude. Disabling this feature causes the beta<br />

counter channels to return a total particle count, including alphas. Users who<br />

wish only to measure total radiation may do so by disabling anti-coincidence and<br />

turning off all alpha count channels through the override parameter screen.<br />

This software algorithm attempts to recognize and compensate for alarm patterns<br />

typically caused by attached radon daughters. Radon compensation is always<br />

disabled when combination alpha/beta emitting isotopes are expected.<br />

Enables/disables the storage of instrument status changes on disk. If enabled, the<br />

latest 200 status changes are stored. Status changes include HIGH<br />

BACKGROUND, OUT OF GAS, COMM FAILURE, HV FAILURE,<br />

CHANNEL SENSITIVITY FAILURE, HIILO COUNT FAILURE, and<br />

NORMAL.<br />

Voice Delay<br />

Store Transactions?<br />

If the voice annunciator option is installed, the <strong>PCM</strong>-2 will delay for this number<br />

of seconds before prompting the user audibly.<br />

A "YES" in this field causes the <strong>PCM</strong>-2 to save measurement results for<br />

transmission to a site host computer. An" ALARMS" in this field causes the<br />

<strong>PCM</strong>-2 to save only alarmed measurement results for transmission to a site host<br />

computer. Note: If this value is set to "YES" or "ALARMS" when the<br />

instrument is not connected to a host computer, the transaction buffer will<br />

eventually overflow, causing the unit to remove itself from service. To reset the<br />

<strong>PCM</strong>2.MANI REV AI April 1995 5-15