Channel Override Parameters<br />

Values for detector parameters may be entered from the override screen as an<br />

alternative to editing individual detectors. Entries made here will be written to all<br />

detectors simultaneously, replacing the existing values. This capability is useful<br />

during initial setup of the instrument, however care must be taken to preserve any<br />

special values required for individual detectors. For explanations of these<br />

parameters, refer to the applicable detector parameter screen documentation.<br />

[HIGH VOLTAGE (VOLTS): 1550<br />

Channel Override Parameters<br />

~Alpha Channels~~~~~~~~~~Beta Channels~~~~~~~~~1<br />

I ~~~~~~~LD Yes<br />

(% MAX): 5<br />

ACTIVE:<br />

THRESHOLD (% MAX):<br />

Yes<br />

45<br />

EFFICIENCY (%): 22.0 EFFICIENCY (%): 22.0<br />

RDA (DPM): 5000 RDA (DPM): 25000<br />

HI FAIL (CPS): 100000 HI FAIL (CPS): 100000<br />

LO FAI L (CPS): 0.00 LO FAIL (CPS): 0.00<br />

WEIGHT FACTOR: 50.0 WEIGHT FACTOR: 10.0<br />

SHIELD FACTOR: 1.00<br />

Instrument Configuration Parameters<br />

~Instrunent Confi gurati on ParameterS===~~~~~~~~~~~~~~~=n<br />

SERIAL NUMBER: 0000 PRINTER TYPE: Epson<br />

COUNT RATE UNITS: CPS USES BEFORE BKG UPDATE: 0<br />

ACTIVITY UNITS: DPM BKG UPDATE AFTER ALARM? No<br />

COUNT MODE: Min Count Time DAY FILE DIRECTORY STORAGE:<br />

10 ENTRY METHOD: Keypad/Aux TRANSACTIONS? No<br />

RWP ENTRY METHOD: None STATUS CHANGES? No<br />

ALARM HOLD (SEC): 1 SOURCE CHECK RESULTS? No<br />

SYSTEM ADDRESS: 2 DISPLAY MIDWAY RESULTS? Yes<br />

SYSTEM BAUD RATE: 9600 HAND SWITCH REQUIRED? Yes<br />

SYSTEM PROTOCOL: uLAN HIP SWITCH REQUIRED? Yes<br />

ANTI-COINCIDENCE: Enabled L.FOOT SWITCH REQUIRED? Yes<br />

RADON COMPENSATION: Disabled R.FOOT SWITCH REQUIRED? Yes<br />

STATUS LOGGING: Disabled SHOULDER SWITCH REQUIRED? No<br />

VOICE DELAY (SEC): 5 ACCESS CONTROL OPTION? No<br />

STORE TRANSACTIONS? No LO LEVEL PASSWORD:<br />

PRINT TRANSACTIONS? No HI LEVEL PASSWORD:<br />

PRINTER INIT:<br />

DAY FILE DIRECTORY: C:\DATA<br />

Serial Number<br />

All reports generated by this <strong>PCM</strong>-2 are identified with this unit number.<br />