WF= Weight Factor<br />

R new = The most-recently measured (real) background count rate,<br />

measured in a single count time interval.<br />

Beta Shield Factor<br />

When the instrument is occupied, some detectors may be shielded from local<br />

radiation sources which account for part of the observed background. The shield<br />

factor is defined as background while occupied divided by background while<br />

unoccupied. Note: Alpha channel shield factors are always equal to 1.<br />

The following values are displayed, but may not be edited:<br />

Alpha & Beta<br />

Background<br />

Alpha & Beta<br />

Alarm Level<br />

Alpha & Beta<br />

Count Time<br />

The current background count rates for this detector.<br />

Net count rates (above background) at which this detector will alarm based on the<br />

current background, RDA, RDA confidence level, sigma factor, etc.<br />

The count times calculated for the alpha and beta channels of this detector given<br />

the current RDA, RDA confidence level, sigma factor and background count<br />

rates.<br />

Hand Probe Parameters<br />

If the hand probe option is installed, this screen may be used to set its high<br />

voltage, alpha and beta thresholds and alpha and beta alarm setpoints. An alarm<br />

setpoint may be entered for both the alpha and beta channels. This setpoint is<br />

expressed in the count rate units selected for the <strong>PCM</strong>-2 (CPS or CPM). In<br />

addition, a weighing factor and a background sigma factor may be entered for<br />

both the alpha and beta channels. These parameters are used in a "smoothing"<br />

algorithm which has been incorporated for the hand probe readings to allow the<br />

user to control the reponse time and the step change for the hand probe readings.<br />

HIGH VOLTAGE (VOLTS): 1550<br />

Hand Probe Parameters<br />

~ALpha Channel<br />

Beta ChanneL<br />

THRESHOLD (% MAX):<br />

ALARM SETPOINT (NCPS)<br />

45<br />

61.6<br />

THRESHOLD (% MAX):<br />

ALARM SETPOINT (NCPS)<br />

5<br />

9.50<br />

Measured VaLue<br />

BACKGROUND (CPS): 0.032<br />

Measured VaLue<br />

BACKGROUND (CPS): 6.19<br />

<strong>PCM</strong>2.MANI REV AI April 1995 5-13