High Voltage<br />

Geometry Factor<br />

Alpha & Beta<br />

Channels Active<br />

Alpha Threshold<br />

Beta Threshold<br />

Alpha & Beta<br />

Efficiencies<br />

Alpha & Beta RDAs<br />

Alpha & Beta<br />

High Fail<br />

Alpha & Beta<br />

Low Fail<br />

Alpha & Beta<br />

Weig ht Factors<br />

The detector voltage may be adjusted in steps of approximately ten volts.<br />

Background count rates are multiplied by this factor before being used to test for<br />

detector sensitivity. This converts average background count rates into count rate<br />

per unit area in order to compensate for differences in detector size. This is used<br />

only for channel sensitivity calculation.<br />

Detector(s) marked as inactive will not be used in any computations. This may be<br />

done to keep the instrument operating if one detector fails.<br />

Particles which cause the detector to produce a pulse exceeding this voltage will<br />

be counted as alpha particles. Lower amplitudes register as betas.<br />

Particles which produce pulses below this amplitude will not be counted as either<br />

alpha or beta.<br />

The ratio of counts per disintegration expressed as a percentage.<br />

Contamination of this activity level will be detected with a probability equal to the<br />

system RDA confidence level.<br />

Detectors which show background levels above these limits will be considered too<br />

noisy or contaminated to use, and will remove the instrument from service.<br />

Detectors with backgrounds below these levels are assumed to have failed. The<br />

instrument will not count if this occurs.<br />

This parameter controls the speed with which the computed average background<br />

rate will follow changes in actual background count rates. The weight factors<br />

smooth statistical fluctuations in background measurements by synthesizing an<br />

exponential moving-average. Increasing the weighting factor will enhance the<br />

smooting function at the expense of extending the time constant associated with<br />

recognizing step changes in the real average background count rate. The formula<br />

for computing the weighted average background count rate is:<br />

fRoid x WF) + R new<br />

R bkg = WF+l<br />

Where: R bkg = The new computed average background count rate<br />

R o1d =<br />

The previous computed average background count rate<br />

5-12<br />

I-'CML.MAN/ Hev A/April I ~~b