Alpha & Beta<br />

Sum Zone Alarm<br />

Alpha & Beta<br />

Sum Channel Alarm<br />

A sum zone is a user defined group of two to four detectors which are grouped<br />

together to enhance the monitor's ability to detect contamination spread over an<br />

area shared by several adjacent detectors. The alpha or beta sum zone alarm<br />

setpoint defines the activity required within the sum zone area before a sum zone<br />

alarm is triggered. Refer to the upcoming section or sum zone setup for more<br />

information on sum zones.<br />

A sum channel is a sum zone consisting of every channel in the monitor. The<br />

alpha or beta sum channel alarm setpoint is the activity which must be present<br />

within the entire monitor before the sum channel alarm is triggered.<br />

Whenever the above parameters are changed, the following dependent values are<br />

recalculated and displayed:<br />

Max Alpha &<br />

Beta Channel RDA<br />

False Alarm Rate<br />

Detector Parameters<br />

At the current background count rates, these are the highest RDA for any channel<br />

in the <strong>PCM</strong>-2 with the specified count time and statistical parameters.<br />

This is the percentage of users who will see false alarms given the current set of<br />

parameters.<br />

Detector parameters are normally set via the Override Parameters<br />

screen. Values entered from this screen are written to all detectors;<br />

any entries made previously to a single detector are lost. Parameters<br />

for a single channel may be examined or adjusted through the Detector<br />

Parameters screen by using the arrow keys to select the desired<br />

detector. Detailed explanations of these parameters and their uses are<br />

provided elsewhere in this manual.<br />

Detector #0 Parameters - Fixed Count Time Mode<br />

~ VOLTAGE (VOLTS), 1600<br />

GEOMETRY FACTOR: 1.00<br />

i"phO Ohooool<br />

I ACTIVE: Yes<br />

I THRESHOLD (% MAX): 45<br />

EFFICIENCY (%): 22.0<br />

I HI<br />

lO FAll (CPS): 100000<br />

FAI l (CPS): 0.00<br />

WEIGHT FACTOR: 50.0<br />

Beta Channel<br />

ACT IVE:<br />

Yes<br />

THRESHOLD (% r'1AX): 5<br />

EFFICIENCY (%): 22.0<br />

HI FAll (CPS): 100000<br />

lO FAll (CPS): 0.00<br />

WEIGHT FACTOR: 10.0<br />

SH IElD FACTOR: 1.00<br />

J<br />

II Calculated Values<br />

IBACKGROUND (CPS):<br />

ALARM lEVEL (NCPS):<br />

RDA lEVEL (DPS):<br />

0.00<br />

1. DO<br />

1. DO<br />

Calculated Values<br />

BACKGROUND (CPS): 0.00<br />

ALARM lEVEL (NCPS): 1. DO<br />

RDA lEVEL (DPS): 1.00<br />

<strong>PCM</strong>2.MANI REV AI April 1995 5-7