High Sensitivity Detector Screens Option (<strong>PCM</strong>2 OPT16)<br />

High Sensitivity<br />

Detector Screens<br />

Option Parts List<br />

These optional 10 mil (0.010 in.) thick, 1/10" etched grid screens serve as a<br />

direct replacement for the standard 0.030 in. thick, 1/8" square perforated plated<br />

steel screens normally furnished with the <strong>PCM</strong>-2 unit for detector face protection<br />

and user body interface. These "high sensitivity" screens, being physically<br />

thinner and having a 20 % greater open area, provide a proportionate decrease in<br />

typical measurement count times when used. They are recommended for usage in<br />

monitoring environments where maximum sensitivity (such as alpha measurement)<br />

is desired. However, they do not provide as smooth of feel or the increased<br />

resistance to wear that the more rugged standard perforated detector screens do.<br />

Reference <strong>PCM</strong>-2 Option 16; High Sensitivity Detector Screens<br />

Item<br />

Ref #<br />

Part Name Description Manufacturer and Part<br />

Number<br />

Eberline Part<br />

Number<br />

010 Detector Screen,<br />

Long<br />

Etched 1!l0"<br />

sq. perf<br />

Eberline Instruments<br />

ZPl1534164<br />

020 Detector Screen,<br />

Medium<br />

Etched 1/10"<br />

sq. perf<br />

Eberline Instruments<br />

ZP1l534058<br />

030 Detector Screen,<br />

Small<br />

Etched 1/10"<br />

sq. perf<br />

Eberline Instruments<br />

ZPl1534090<br />

Head & Shoulders Position Sensor Option (<strong>PCM</strong>2 OPT17)<br />

This option consists of a small photocell emitter/receiver pair which are mounted<br />

to the upper front detector bank of the <strong>PCM</strong>-2 unit for the purpose of sensing the<br />

position of the user's upper body relative to the subject detectors. The supporting<br />

electronics for the photocell pair are mounted within the upper electronics<br />

enclosure of the instrument. The photo-emitter resides in a small stainless steel<br />

housing which front mounts at lower center right of the upper main detector panel<br />

betweeen the right-hand and center detectors. Its light beam traverses the center<br />

detector upwardly aligned with the photoreceiver which is mounted directly above<br />

a small hole in the ceiling of the unit adjacent to the head detector. The light<br />

beam is spaced approximately 0.25 in. away from the frontal surface of the center<br />

detector on the front panel. The angular alignment of the photocell's beam and<br />

17-90 <strong>PCM</strong>2.MAN/ Rev A/April 1995