1400 to 1900 volts) using long count times (at least 300 seconds per voltage).<br />

Once the test has been started, no operator intervention is required; it is therefore<br />

advised that this plateau be nm overnight.<br />

Once the background measurements have been completed, a plateau screen is<br />

available for each detector. Page through these displays and verify that all<br />

detectors have similar voltage curves. Note the voltage range over which alpha<br />

background rates begin to increase; use the left and right arrow keys to move the<br />

vertical cursor line and note that the exact count rates displayed at the bottom of<br />

the screen are displayed for each voltage. Best results will be obtained at a<br />

voltage slightly below the point at which alpha background rates begin to increase.<br />

The SOURCE PLATEAU program determines the response of anyone detector to<br />

alpha and beta sources. Select starting and ending voltages which bracket the<br />

operating point determined from the background plateau results and count times<br />

appropriate to the sources to be used. If large sources are available, it should not<br />

be necessary to correct for background rates. If a background plateau has already<br />

been run using the same voltage increments and covering a range which matches<br />

the source plateau high voltage range, background corrections may be applied.<br />

Both alpha and beta source plateaus should be run to obtain a full set of results.<br />

Once an operating voltage has been determined, enter it through the OVERRIDE<br />

PARAMETERS screen to set all detectors to this value. Individual detector<br />

voltages may be adjusted through the DETECTOR PARAMETERS screen or the<br />

VIEW PLATEAU display for that channel. Return to the BACKGROUND<br />

AVERAGES screen and verify that all detectors give reasonable count rates at the<br />

selected voltage. It is recommended that source plateaus be run on at least two<br />

detectors to insure that the operating voltage selected will be appropriate for the<br />

entire system. Verify similar results for all detectors that are plateaued.<br />

Detector Efficiency<br />

Calibration<br />

Using the EFFICIENCIES conunand under the CALIBRATION heading, each<br />

detector must be calibrated with alpha and beta sources of known activities. 100<br />

square em "plate" sources are preferred for this purpose as opposed to smaller<br />

"coin" sources. High-activity sources should be used if possible, since they will<br />

provide accurate results with short count times.<br />

Enter the activities of the sources to be used (the instrument will correct for<br />

source decay if given an initial calibration date and the isotope half-life). Follow<br />

the instructions displayed to select a detector; position the source on that detector<br />

and begin the count. Once a satisfactory efficiency measurement has been<br />

completed, store that value and move the source to the next detector. After all<br />

detectors have been calibrated with one source (alpha or beta), repeat the<br />

procedure with the other after selecting the correct channel type.<br />

<strong>PCM</strong>2.MANI REV AI April 1995 2-7