The RDA confidence should be edited if it is to be other than the default<br />

value of 95 %.<br />

The count time (max count time in Mode 3) is set in this screen.<br />

Edit Instrument Parameters<br />

The preferred count rate units (cpm or cps), activity units (dpm, dps, Bq, or<br />

nCi), and count mode are set in this screen. The three counting modes are:<br />

MODE 1 - Preset All<br />

MODE 2 - Fixed Count Time (Maximum Sensitivity)<br />

MODE 3 - Minimum Count Time<br />

If reporting or logging of data either locally or over a network installation is<br />

desired, additional parameters will need to be configured in this screen.<br />

Refer to section 5 for more information on configuration parameters.<br />

Quick-Start Instructions<br />

These instructions are intended to assist the first-time user in setting up Eberline<br />

<strong>PCM</strong>-2 systems. Once this initial setup has been completed, adjustments may be<br />

made to suit site-specific requirements and policies. Some familiarity with<br />

instruments using gas proportional radiation detectors is assumed in these instructions.<br />

Refer to the technical sections of this manual for detailed information regarding the<br />

unit and its operation.<br />

Power Connections<br />

Counting Gas<br />

Supply<br />

The <strong>PCM</strong>-2 may be operated from either 90-132 or 180-264 Volts, 50/60 Hz.<br />

Voltage range selection is made by a slide switch located on the computer power<br />

supply inside the electronics enclosure. Power is brought in via a line cord<br />

attached to the outlet strip located inside the unit. Remove either of the two cover<br />

plates located on the top of the unit and at the lower rear corner and route this<br />

cord out to a convenient power receptacle. Insure that the switch on the power<br />

strip is on, then turn on the unit by actuating the power switch on the side of the<br />

electronics enclosure. Normal operation is indicated by fan noise from the<br />

computer power supply. Within a few seconds, the disk drive will also begin to<br />

operate, lights on the detector boards will flash, and text will appear on the CRT.<br />

Two gas inlet hoses exit from the lower right corner of the electronics enclosure.<br />

Each of these should be connected to a supply of P-lO gas (10% methane, 90%<br />

argon) at a regulated pressure of 5 psi. Insure that both gas supplies are regulated<br />

to the same pressure so that flow rates will not change when the alternate source<br />

<strong>PCM</strong>2.MI\N/ REV I\/I\pril 1995 2-G