Other Optional Connections<br />

= ."=~~;,£==~-<br />

Eberline offers as options; a report printer, a remote status and alarm indicator,<br />

and a variety of data interfaces to connect the <strong>PCM</strong>-2 to a host computer. Each<br />

of these options is supplied with the required hardware, including cables, and with<br />

specific installation information. As with the power cord, these cables may exit<br />

through either power inlet plate; at the top of the unit or the lower rear corner of<br />

the cabinet.<br />

Configuration<br />

After the instrument has been properly installed and purged with counting gas,<br />

configuration with the correct operating parameters must be performed before it is<br />

placed into service. Configuration is soft-set through the test mode edit routines<br />

via keyboard input. The configuration routines and the parameters of interest<br />

include at the minimum:<br />

Edit Detector Parameters / Edit Override Parameters<br />

NOTE: Detector parameters can be set for individual detectors using the<br />

Edit Detector Parameter routine. The Edit Override Parameters is used to<br />

set the same parameters for all 34 detectors globally.<br />

Detector high voltage and efficiencies should be determined by calibration<br />

routines before being set. The <strong>PCM</strong>-2 is fully calibrated at the factory<br />

which is located at an elevation of 6,480 ft. above sea level. If the<br />

instrument is installed at a different elevation it will be necessary to replateau<br />

to determine the correct high voltage setting. Adjusting the high<br />

voltage setting will necessitate measuring new efficiencies. Reference<br />

Section 7 for detailed information on high voltage settings and detector<br />

calibration.<br />

RDA (Reliably Detectable Activity) levels should be set to the values<br />

required by plant administration. RDA levels can be edited in Detector<br />

Parameters, Override Parameters, or System Parameters.<br />

Edit System Parameters<br />

As noted above, RDA levels can be set for all channels in this screen. If<br />

Mode 2 is being used, this parameter is RDA UPPER LIMIT. Default<br />

values are 5,000 dpm for the beta channels and 900 dpm for the alpha<br />

channels.<br />

2-4<br />

<strong>PCM</strong>2.M/\N/ Rev /\//\pril 1995