Electrical Power ~nsta~<br />

Each of the sixteen detector chambers is connected to two gas plumbing<br />

harnesses/manifolds; A red gas supply harness and a blue gas exhaust harness.<br />

For faster purging, the supply hose should always connect to the lower-most gas<br />

fittings of vertically mounted detectors.<br />

If the Hand Probe Option (Option 1) is installed, a third gas flow meter and a<br />

separate rate adjustment needle valve is provided. This single hand-held detector<br />

will operate adequately on 5-10 cc/minute of PIO counting gas regardless of<br />

whether or not a gas manager is present.<br />

For units equipped with Spare Purging Detectors (Option 15), no special gas flow<br />

rate adjustment considerations need be taken as these spare detectors are provided<br />

with exhaust gas downstream of the main detectors.<br />

For units equipped '';I'ith a Gas Manager (Option 12), refer to the Options Section<br />

for detailed system adjustment direction as operation of the gas supply/exhaust<br />

system is signifiacntly different from that of a standard unit.<br />

Power is brought into the <strong>PCM</strong>-2 via a four-outlet power strip. Into this are<br />

plugged power cords from the unit's lower electronics enclosure and options such<br />

as a printer or badge reader. The line cord attached to this power strip may exit<br />

the unit either through the lower rear corner of the cabinet or through the top of<br />

the unit near its right-hand side, per user request. The default configuration for<br />

power cord egress is through the lower rear power inlet plate.<br />

While the <strong>PCM</strong>-2 is relatively immune to line noise and transients, it is possible<br />

for very large power line spikes to interfere with the instrument. Avoid<br />

connecting the <strong>PCM</strong>-2 to power circuits shared VJith large motors or other<br />

inductive loads. If clean line power is not available, a line conditioner should be<br />

installed.<br />

To prevent unauthorized operation, the unit's on/off switch is placed on the right<br />

side of the electronics enclosure internal to the unit behind the side door. This is<br />

also the location of the fuse holder. If the fuse must be replaced, use a 3-Amp<br />

1.25 x 0.25 inch slow blow fuse rated for 250 Volts.<br />

Two operating voltage ranges are supported; 90-132 and 180-264 volts. This<br />

selection is made by moving a recessed slide switch on the main computer power<br />

supply inside the lower electronics enclosure. An inspection mirror may be used<br />

to verify the position of this switch, which is located between the two outlets on<br />

the right side of the power supply.<br />

<strong>PCM</strong>2.MANI REV AI April 1995 2-3