08.06.2015 Views

PCM-2 Manual.pdf - Voss Associates

PCM-2 Manual.pdf - Voss Associates

PCM-2 Manual.pdf - Voss Associates

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Counting Gas Installation<br />

'~_==~~='-="_.===" ='m===~==· =._==~~~" _~~=~_~=<br />

_<br />

The <strong>PCM</strong>-2 uses gas proportional detectors which require an approximate total of<br />

200 cc/rnin of P-lO counting gas for their operation. Two gas inputs are<br />

provided, for which supply hose feed through grommets at the rear side columns<br />

of the instrument, are provided. The instrument will select one gas source upon<br />

power-up and continue to use that source until its pressure fails, at which time the<br />

alternate gas source will automatically be selected via internal gas switching<br />

electronics. If both inputs drop below approximately 2 psi, the instrument will<br />

cease to operate.<br />

If portable cylinders are used to supply counting gas, each must be regulated<br />

down to approximately 5 psi and connected to one of the gas inputs. The<br />

standard <strong>PCM</strong>-2 unit includes one (1) gas bottle pressure regulator located in the<br />

extraneous parts box provided, as previously noted. If the optional Gas Bottle<br />

Enclosure (Option 9) is ordered, one (1) additional gas pressure regulator is<br />

included. If two counting gas sources are installed it is important that both<br />

regulators be set to the same pressure so that gas flow will remain constant when<br />

the second cylinder is selected. If the pressure at either input fails, a message will<br />

be displayed on the CRT indicating that the gas cylinder needs to be replaced.<br />

When the instrument is operated from a permanently installed counting gas<br />

manifold, both inputs may be connected to the same 5 psi source. This will<br />

eliminate false "gas bottle empty" messages.<br />

Gas Flow<br />

Adjustment<br />

Two flow meters are provided, along with a single needle valve to adjust the gas<br />

flow rate. For normal operation, use the low-range meter to set a flow rate of<br />

approximately 200 cc/minute. When setting up a new instrument or an instrument<br />

which has been disconnected from its gas supply, adjust the flow to 0.8 liter per<br />

minute (800 cc/min) for four hours to purge the detectors completely. Do not<br />

exceed 1 liter per minute (1000 cc/miu) at any time to avoid bursting the<br />

unit's detectors. If replacing a detector, the replacement detector should first be<br />

purged on the bench prior to installaton or purged in place after installation using<br />

the "Quick Detector Purge" gas supply hose coil located behind the flowmeter<br />

mounting bracket. If the latter method is used for detector purging, simply<br />

disconnect the red gas supply hose from the subject detector and plug the quick<br />

purge line in its place. This effectively re-routes the gas flow from the system's<br />

gas supply harness to the purging detector only. Utilization of this quick purging<br />

hose feature will provide complete purging of a large <strong>PCM</strong>-2 detector within<br />

approximately 20 minutes at a nominal gas flow rate of 200cc/min. Faster puging<br />

times can be achieved by increasing the gas flow rate, however, caution must be<br />

exercised to avoid bursting the detector's mylar face. A more thourough<br />

explainatiol1 of the quick purging line feature appears in the maintenance section<br />

(Section 9) of this manual.<br />

2-2<br />

rCM2.MI\N/ Hcv I\/I\pril 1995

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!