Wöhler E 98 Econometer<br />
Measuring Instruments<br />
O 2<br />
CO 2<br />
E 98 Econometer<br />
COInd<br />
350ppm<br />
COv<br />
3000ppm<br />
T FlueGas<br />
T Air<br />
T Dewp.<br />
Draft<br />
100 Pa<br />
DP<br />
40 hPA<br />
l<br />
η<br />
qA<br />
E 98 Maxi Econometer<br />
IR<br />
TÜV approved<br />
The E 98 Econometer - high performance in a compact size ...<br />
Take advantage of the E 98’s performance vs. cost and equip each of your technicians with an E 98 Econometer.<br />
For the heating business this new compact device for flue-gas analysis is the perfect way to save time and money.<br />
With a single click on the button of the E 98 it shows you the flue-gas components and combustion efficiency on<br />
the display.<br />
Application:<br />
• Adjustments on oil and gas-fired<br />
heating systems.<br />
• Replaces old-fashioned analyzers and<br />
does much more.<br />
Advantages:<br />
• data memory for 10 sets<br />
of records and IR interface for<br />
printing and continuous data<br />
transmission<br />
• optional 1.7 m hose (E 98 Maxi)<br />
• additional pressure connection for<br />
gas pressure adjustments or<br />
differential pressure<br />
measurements<br />
• quick response time<br />
• uses regular or rechargeable<br />
batteries<br />
• long operating time: e.g. 38<br />
hours of operation with Mignon<br />
dry cells (2.3 Ah) or 10 hours<br />
with standard NiCd rechargeables<br />
(600 mAh)<br />
• easy to drain and clean water<br />
trap<br />
1. Switch on E98:<br />
2. Select fuel type<br />
3. Sample<br />
4. Let calculate<br />
5. Printout<br />
O 2<br />
measurement<br />
Sensor: electrochemical<br />
Range: 0...20.9 Vol.-%<br />
Accuracy: 0.3 Vol.-% absolute<br />
CO indicator (not H 2<br />
compensated)<br />
Sensor: Semiconductor<br />
Range: 0...350 ppm<br />
Accuracy: 50 ppm<br />
CO sensor<br />
Sensor:<br />
Range:<br />
Accuracy:<br />
(optional)<br />
electrochemical<br />
0...3,000 ppm<br />
± 20 ppm up to 400 pm,<br />
rest 5 % f.m.<br />
Ambient air temperature<br />
measurement<br />
Range: 0...99.9 °C with T F<br />
probe<br />
Accuracy: 2 °C: 0...50 °C<br />
opt. (for measurements accord. to pollution<br />
control regulations)<br />
Range: -20.0...99.9 °C with TA<br />
probe<br />
Accuracy: 1 °C: 0...50 °C<br />
Stack gas temperature measurement<br />
Range: 0...800 °C<br />
Accuracy: 2 °C: 0...125 °C<br />
3 °C: 125...250 °C<br />
4 °C: 250...400 °C<br />
8 °C: >400°C<br />
Differential pressure<br />
Range:<br />
Accuracy:<br />
-100...+4,000 Pa<br />
3 Pa: -20...100 Pa<br />
3 % FSD: -1.0...4.0 hPa<br />
CO 2<br />
determination<br />
Range:<br />
0...CO 2max<br />
Resolution: 0.1 %<br />
Heat loss determination<br />
Depending on fuel, resolution 0.1 %<br />
Combustion efficiency<br />
Range: 0...120.0 %<br />
Resolution: 0.1 %<br />
Dewpoint temperature measurement<br />
Depending on fuel, resolution 0.1 °C<br />
Condensate quantity<br />
Display in kg/m (gas) or kg/l (oil)<br />
12<br />
Internet: http://www.woehler.com